Application of commercial MOSFET detectors forin vivodosimetry in the therapeutic x-ray range from 80 kV to 250 kV
Authors:
- Christian EhringfeldCEChristian Ehringfeld
- Susanne SchmidSSSusanne Schmid
- Karin PoljancKPKarin Poljanc
- Christian KirisitsChristian Kirisits
- Hannes AigingerHAHannes Aiginger
- Dietmar GeorgDietmar Georg
- 7 January 2005
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 50 (2) , 289-303
Abstract
The purpose of this study was to investigate the dosimetric characteristics (energy dependence, linearity, fading, reproducibility, etc) of MOSFET detectors for in vivo dosimetry in the kV x-ray range. The experience of MOSFET in vivo dosimetry in a pre-clinical study using the Alderson phantom and in clinical practice is also reported. All measurements were performed with a Gulmay D3300 kV unit and TN-502RDI MOSFET detectors. For the determination of correction factors different solid phantoms and a calibrated Farmer-type chamber were used. The MOSFET signal was linear with applied dose in the range from 0.2 to 2 Gy for all energies. Due to fading it is recommended to read the MOSFET signal during the first 15 min after irradiation. For long time intervals between irradiation and readout the fading can vary largely with the detector. The temperature dependence of the detector signal was small (0.3% degrees C(-1)) in the temperature range between 22 and 40 degrees C. The variation of the measuring signal with beam incidence amounts to +/-5% and should be considered in clinical applications. Finally, for entrance dose measurements energy-dependent calibration factors, correction factors for field size and irradiated cable length were applied. The overall accuracy, for all measurements, was dominated by reproducibility as a function of applied dose. During the pre-clinical in vivo study, the agreement between MOSFET and TLD measurements was well within 3%. The results of MOSFET measurements, to determine the dosimetric characteristics as well as clinical applications, showed that MOSFET detectors are suitable for in vivo dosimetry in the kV range. However, some energy-dependent dosimetry effects need to be considered and corrected for. Due to reproducibility effects at low dose levels accurate in vivo measurements are only possible if the applied dose is equal to or larger than 2 Gy.Keywords
This publication has 16 references indexed in Scilit:
- Characterization of high-sensitivity metal oxide semiconductor field effect transistor dosimeters system and LiF:Mg,Cu,P thermoluminescence dosimeters for use in diagnostic radiologyApplied Radiation and Isotopes, 2002
- Build-up modification of commercial diodes for entrance dose measurements in ‘higher energy’ photon beamsRadiotherapy and Oncology, 1999
- In vivo dosimetry during external photon beam radiotherapyInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1999
- The Characterization of a Commercial Mosfet Dosimeter System for Use in Diagnostic X-RayHealth Physics, 1998
- Quality assurance in radiation oncology. A study of feasibility and impact on action levels of an in vivo dosimetry program during breast cancer irradiationRadiotherapy and Oncology, 1998
- The response of a MOSFET, p-type semiconductor and LiF TLD to quasi-monoenergetic x-raysPhysics in Medicine & Biology, 1997
- A new radiotherapy surface dose detector: The MOSFETMedical Physics, 1996
- A miniature MOSFET radiation dosimeter probeMedical Physics, 1994
- A solid water phantom material for radiotherapy x-ray and γ-ray beam calibrationsMedical Physics, 1982
- The theoretical and microdosimetric basis of thermoluminescence and applications to dosimetryPhysics in Medicine & Biology, 1981