Meta‐analysis of complications in aortic valve replacement: Comparison of Medtronic‐Corevalve, Edwards‐Sapien and surgical aortic valve replacement in 8,536 patients
Authors:
- 13 March 2012
- journal article
- research article
- Published by Wiley in Catheterization and Cardiovascular Interventions
- Vol. 80 (1) , 128-138
Abstract
Background: We sought to establish the complication rates following transcatheter aortic valve replacement (TAVR) in the context of high risk and octogenarian surgical aortic valve replacement (SAVR) in the contemporary literature, and to critically analyze population characteristics and outcomes. Methods: TAVR studies were selected from nonoverlapping series and SAVR studies for comparison if they met similar entry criteria. Bayesian meta‐analytic methods were employed. Results: For the 5024 TAVR and 3512 SAVR patients included in the study, TAVR subjects had greater baseline renal impairment (P < 0.001), a higher incidence of prior myocardial infarction (P = 0.032) and respiratory disease (P = 0.005) and a higher logistic EuroSCORE (P = 0.039). There were no significant differences observed in complications studied in SAVR and TAVR: 30 day mortality (9% vs 8.5%, P = 0.31), 1 year mortality (18.4% vs 22.8%, P = 0.65), 30 day stroke (2.4% vs 2.6%, P = 0.72), new permanent pacemaker (5.9% vs 12.1%, P = 0.055) and dialysis inception (2.4% vs 4.1%, P = 0.70). We also compared demographics and outcomes between the two types of transcatheter valves. Apart from some variation in functional status, there were no significant differences at baseline with different TAVR designs. The only difference in complications was the need for pacemaker insertion, higher with the Medtronic‐Corevalve than with the Edwards‐Sapien design (24.5% vs 5.9% P < 0.0001). Conclusions: Complications for elderly and high risk aortic stenosis patients being treated by TAVR appear comparable to those selected for SAVR in the real‐world.Keywords
This publication has 45 references indexed in Scilit:
- Association of aortic valve calcification severity with the degree of aortic regurgitation after transcatheter aortic valve implantationInternational Journal of Cardiology, 2011
- Safety and Efficacy of the Subclavian Approach for Transcatheter Aortic Valve Implantation With the CoreValve Revalving SystemCirculation: Cardiovascular Interventions, 2010
- Bayesian Methods for Data AnalysisAmerican Journal of Ophthalmology, 2010
- Relationship between the logistic EuroSCORE and the Society of Thoracic Surgeons Predicted Risk of Mortality score in patients implanted with the CoreValve ReValving System—A Bern-Rotterdam StudyAmerican Heart Journal, 2010
- Paradoxical low flow and/or low gradient severe aortic stenosis despite preserved left ventricular ejection fraction: implications for diagnosis and treatmentEuropean Heart Journal, 2009
- Frailty after aortic valve replacement (AVR) in octogenariansArchives of Gerontology and Geriatrics, 2009
- Using BayesianP-Values in a 2 × 2 Table of Matched Pairs with Incompletely Classified DataJournal of the Royal Statistical Society Series C: Applied Statistics, 2009
- Percutaneous Aortic Valve Replacement Will Become a Common Treatment for Aortic Valve DiseaseJACC: Cardiovascular Interventions, 2008
- Aortic valve replacement in octogenariansJournal of Cardiothoracic Surgery, 2007
- Aortic Valve Replacement and Concomitant Coronary Artery Bypass: Assessing the Impact of Multiple GraftsThe Annals of Thoracic Surgery, 2007