Dynamic Biomechanical Examination of the Lumbar Spine with Implanted Total Spinal Segment Replacement (TSSR) Utilizing a Pendulum Testing System
Authors:
Open Access
- 25 February 2013
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 8 (2) , e57412
Abstract
Biomechanical investigations of spinal motion preserving implants help in the understanding of their in vivo behavior. In this study, we hypothesized that the lumbar spine with implanted total spinal segment replacement (TSSR) would exhibit decreased dynamic stiffness and more rapid energy absorption compared to native functional spinal units under simulated physiologic motion when tested with the pendulum system. Five unembalmed, frozen human lumbar functional spinal units were tested on the pendulum system with axial compressive loads of 181 N, 282 N, 385 N, and 488 N before and after Flexuspine total spinal segment replacement implantation. Testing in flexion, extension, and lateral bending began by rotating the pendulum to 5°; resulting in unconstrained oscillatory motion. The number of rotations to equilibrium was recorded and bending stiffness (N-m/°) was calculated and compared for each testing mode. The total spinal segment replacement reached equilibrium with significantly fewer cycles to equilibrium compared to the intact functional spinal unit at all loads in flexion (p<0.011), and at loads of 385 N and 488 N in lateral bending (p<0.020). Mean bending stiffness in flexion, extension, and lateral bending increased with increasing load for both the intact functional spinal unit and total spinal segment replacement constructs (p0.18). Lumbar functional spinal units with implanted total spinal segment replacement were found to have similar dynamic bending stiffness, but absorbed energy at a more rapid rate than intact functional spinal units during cyclic loading with an unconstrained pendulum system. Although the effects on clinical performance of motion preserving devices is not fully known, these results provide further insight into the biomechanical behavior of this device under approximated physiologic loading conditions.Keywords
This publication has 35 references indexed in Scilit:
- Optimal stiffness of a pedicle-screw-based motion preservation implant for the lumbar spineEuropean Spine Journal, 2011
- Does anterior lumbar interbody fusion promote adjacent degeneration in degenerative disc disease? A finite element studyJournal of Orthopaedic Science, 2011
- Effect of multilevel lumbar disc arthroplasty on spine kinematics and facet joint loads in flexion and extension: a finite element analysisEuropean Spine Journal, 2010
- In Situ Contact Analysis of the Prosthesis Components of Prodisc-L in Lumbar Spine Following Total Disc ReplacementSpine, 2009
- Motion Preservation Surgery of the Spine: Advanced Techniques and ControversiesAmerican Journal of Neuroradiology, 2009
- Biomechanical Effect of Constraint in Lumbar Total Disc ReplacementSpine, 2009
- Effect of an artificial disc on lumbar spine biomechanics: a probabilistic finite element studyEuropean Spine Journal, 2008
- Early stage disc degeneration does not have an appreciable affect on stiffness and load transfer following vertebroplasty and kyphoplastyEuropean Spine Journal, 2008
- The dynamic flexion/extension properties of the lumbar spine in vitro using a novel pendulum systemJournal of Biomechanics, 2007
- Investigation of Vibration Characteristics of the Ligamentous Lumbar Spine Using the Finite Element ApproachJournal of Biomechanical Engineering, 1994