Comparison of 4 methods for quantifying posterior capsule opacification
Authors:
- Wolf BuehlWBWolf Buehl
- Rupert MenapaceRMRupert Menapace
- Michael GeorgopoulosMGMichael Georgopoulos
- Georg RainerGRGeorg Rainer
- Hannes SieglHSHannes Siegl
- Alexandra KaiderAKAlexandra Kaider
- Axel PinzAPAxel Pinz
- 1 January 2003
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 29 (1) , 106-111
Abstract
To compare the results of posterior capsule opacification (PCO) quantification and the repeatability of a fully automated analysis system (Automated Quantification of After-Cataract [AQUA]) with that of 2 other quantification methods and subjective grading of PCO. A test set of digital retroillumination images of 100 eyes with PCO of varying degrees was used. Department of Ophthalmology, University of Vienna, Vienna, Austria. One hundred digital retroillumination images of eyes (100 patients) with PCO were selected to attain an even distribution from mild to severe cases. The images were evaluated by 4 methods: subjective grading by 4 experienced and 4 inexperienced examiners, the subjective Evaluation of Posterior Capsular Opacification (EPCO) system, posterior capsule opacification (POCO) software, and the AQUA system. Ten images were presented twice to assess the reproducibility of the analysis systems. Subjective grading correlated best with the subjective EPCO system and the objective AQUA system (r = 0.94 and r = 0.93, respectively). The POCO system showed very early saturation and therefore a much weaker correlation (r = 0.73). The POCO scores reached the maximum of 100% in several minimal to mild PCO cases. The reproducibility of the AQUA software was perfect and that of the other analysis systems, comparably satisfactory. The objective AQUA score correlated well with subjective methods including the EPCO system. The POCO system, which assesses PCO area, did not adequately describe PCO intensity and includes a subjective step in the analysis process. The AQUA system could become an important tool for randomized masked trials of PCO inhibition.Keywords
This publication has 7 references indexed in Scilit:
- Reproducibility of standardized retroillumination photography for quantification of posterior capsule opacificationJournal of Cataract & Refractive Surgery, 2002
- Posterior capsule opacificationJournal of Cataract & Refractive Surgery, 1999
- A systematic overview of the incidence of posterior capsule opacificationOphthalmology, 1998
- Relationship between intraocular lens biomaterials and posterior capsule opacificationJournal of Cataract & Refractive Surgery, 1998
- Photographic image analysis system of posterior capsule opacificationJournal of Cataract & Refractive Surgery, 1997
- High-resolution digital retroillumination imaging of the posterior lens capsule after cataract surgeryJournal of Cataract & Refractive Surgery, 1997
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986