An x−ray spectrometer for laser−induced plasmas
Authors:
- 1 January 1975
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 46 (1) , 53-57
Abstract
A seven−channel K−edge filter spectrometer which uses Si PIN−type diodes has been developed as an x−ray diagnostic tool for laser plasma experiments. The spectrometer is designed to sample the photon energy range from 500 eV to 20 keV. The theory and techniques involved in the application of a K−edge filter−Si diode spectrometer to laser plasma studies are presented. Electron temperatures in plasmas are often determined by the two−foil ratio method. The present work defines conditions under which such a measurement may prove unreliable. Typical absolute x−ray spectra obtained from subnanosecond irradiation of 1000 μ CD2 spheres with a few joules of laser energy yields 200 eV electron temperatures and a 1% laser to x−ray conversion efficiency.Keywords
This publication has 5 references indexed in Scilit:
- An x-ray spectral measurement system for nanosecond plasmasReview of Scientific Instruments, 1974
- Simple X-Ray Bandpass Filters for Nanosecond Resolution of Plasma SpectraReview of Scientific Instruments, 1972
- Production and Diagnostic Measurements of Kilovolt High-Density Deuterium, Helium, and Neon PlasmasJournal of Applied Physics, 1966
- Semiconductor detectors for nuclear spectrometry, INuclear Instruments and Methods, 1966
- Continuum Radiation in the X Ray and Visible Regions from a Magnetically Compressed Plasma (Scylla)Physical Review B, 1960