Electrical property sensing biopsy needle for prostate cancer detection
Authors:
- 1 June 2013
- journal article
- review article
- Published by Wiley in The Prostate
- Vol. 73 (15) , 1603-1613
Abstract
BACKGROUND Significant electrical property differences have been demonstrated to exist between malignant and benign prostate tissues. We evaluated how well a custom designed clinically deployable electrical property sensing biopsy needle is able to discriminate between these tissue types in an ex vivo prostate model. METHODS An electrical impedance spectroscopy (EIS) sensing biopsy (Bx) needle was developed to record resistive (ρR) and reactive (ρX) components of electrical impedance from 100 Hz to 1 MHz. Standard twelve-core biopsy protocols were followed, in which the EIS-Bx device was used to gauge electrical properties prior to extracting tissue cores through biopsy needle firing from 36 ex vivo human prostates. Histopathological assessment of the cores was statistically compared to the impedance spectrum gauged from each core. RESULTS The magnitudes of the mean resistive and reactive components were significantly higher in cancer tissues (P < 0.05). ROC curves showed that ρR at 63.09 kHz was optimal for discriminating cancer from benign tissues; this parameter had 75.4% specificity, 76.1% sensitivity, and ROC AUC of 0.779. Similarly, 251.1 kHz was optimal when using ρX to discriminate cancer from benign tissues; this parameter had a 77.9% specificity, 71.4% sensitivity, and ROC AUC of 0.79. CONCLUSION Significant electrical property differences noted between benign and malignant prostate tissues suggest the potential efficacy an EIS-Bx device would provide for cancer detection in a clinical setting. By sensing a greater fraction of the prostate's volume in real-time, the EIS-Bx device has the potential to improve the accuracy of cancer grading and volume estimation made with current biopsy procedures. Prostate 73: 1603–1613, 2013.Keywords
This publication has 22 references indexed in Scilit:
- Screening for Prostate Cancer: U.S. Preventive Services Task Force Recommendation StatementAnnals of Internal Medicine, 2012
- Overview of Dynamic Contrast-Enhanced MRI in Prostate Cancer Diagnosis and ManagementAmerican Journal of Roentgenology, 2012
- Imaging in Prostate Cancer Staging: Present Role and Future PerspectivesUrologia Internationalis, 2012
- Diffusion-weighted Endorectal MR Imaging at 3 T for Prostate Cancer: Tumor Detection and Assessment of AggressivenessRadiology, 2011
- Radical Prostatectomy versus Watchful Waiting in Early Prostate CancerNew England Journal of Medicine, 2011
- Functional Imaging of Localized Prostate Cancer Aggressiveness Using 11C-Acetate PET/CT and 1H-MR SpectroscopyJournal of Nuclear Medicine, 2010
- New horizons in prostate cancer imagingEuropean Journal of Radiology, 2009
- Combined Magnetic Resonance Imaging and Magnetic Resonance Spectroscopy Imaging in the Diagnosis of Prostate Cancer: A Systematic Review and Meta-analysisEuropean Urology, 2009
- Extended and Saturation Prostatic Biopsy in the Diagnosis and Characterisation of Prostate Cancer: A Critical Analysis of the LiteratureEuropean Urology, 2007
- Bioimpedance: Novel use of a minimally invasive technique for cancer localization in the intact prostateThe Prostate, 1999