Diagnosis of arterial occlusive disease of the upper extremities: Comparison of color duplex sonography and angiography
Authors:
- 29 September 2003
- journal article
- research article
- Published by Wiley in Journal of Clinical Ultrasound
- Vol. 31 (8) , 407-411
Abstract
Purpose: The aim of this study was to assess the abilities of color duplex sonography (CDUS) to detect and characterize arterial occlusive disease of the upper extremities.Methods: We prospectively compared the results of CDUS with those of intra‐arterial digital subtraction angiography, which were considered definitive, in patients with symptomatic arterial occlusive disease of the upper extremities. In each extremity, we visualized 9 arterial segments, which were each evaluated for stenoses and occlusions. Each segment was categorized on each imaging modality as not significantly narrowed (narrowed by <50%), significantly narrowed (narrowed by ≥50% but <100%), or occluded (100%).Results: We examined a total of 578 segments in 57 patients (34 men and 23 women) with a mean age of 50 years (range, 20–74 years). CDUS had a sensitivity, a specificity, a positive predictive value, a negative predictive value, and an accuracy of 98%, 99%, 97%, 99.5%, and 99%, respectively, for detecting occluded lesions and 79%, 100%, 100%, 99%, and 99%, respectively, for detecting hemodynamically significantly stenotic lesions. Notably, the sensitivity of CDUS for diagnosing significantly stenotic lesions (79%) was lower than that for diagnosing occlusive disease (98%).Conclusions: With high sensitivity and accuracy rates, CDUS is a reliable screening method for detecting arterial occlusive disease of the upper extremities. This modality efficiently provides anatomic and hemodynamic data that are useful in cases of such disease. © 2003 Wiley Periodicals, Inc. J Clin Ultrasound 31:407–411, 2003Keywords
This publication has 10 references indexed in Scilit:
- Diagnosis of arterial disease of the lower extremities with duplex ultrasonographyBritish Journal of Surgery, 1996
- Limitations of ultrasonic duplex scanning for diagnosing lower limb arterial stenoses in the presence of adjacent segment diseaseJournal of Vascular Surgery, 1994
- Arterial sonography: efficacy for the diagnosis of arterial disease of the lower extremity.American Journal of Roentgenology, 1993
- Lower limb colour flow imaging: A comparison with ankle: Brachial measurements and angiographyClinical Radiology, 1993
- Color flow doppler ultrasonography: Comparison with peripheral arteriography for the investigation of peripheral vascular diseaseJournal of Clinical Ultrasound, 1992
- Accuracy of lower extremity arterial duplex mappingJournal of Vascular Surgery, 1992
- Duplex scanning of the peripheral arteries: Correlation of the peak velocity ratio with angiographic diameter reductionUltrasound in Medicine & Biology, 1992
- Peripheral arterial occlusive disease: prospective comparison of MR angiography and color duplex US with conventional angiography.Radiology, 1991
- Determination of the extent of lower-extremity peripheral arterial disease with color-assisted duplex sonography: comparison with angiography.American Journal of Roentgenology, 1990