A fast cardiac gamma camera with dynamic SPECT capabilities: design, system validation and future potential
Authors:
Open Access
- 29 June 2010
- journal article
- research article
- Published by Springer Nature in European Journal of Nuclear Medicine and Molecular Imaging
- Vol. 37 (10) , 1887-1902
Abstract
Purpose The goal of this study is to present the Discovery NM 530c (DNM), a cardiac SPECT camera, interfacing multi-pinhole collimators with solid-state modules, aiming at slashing acquisition time without jeopardizing quality. DNM resembles PET since it enables 3-D SPECT without detector motion. We further envision how these novel capabilities may help with current and future challenges of cardiac imaging. Methods DNM sensitivity, spatial resolution (SR) and energy resolution (ER), count rate response, cardiac uniformity and cardiac defect contrast were measured and compared to a dedicated cardiac, dual-head standard SPECT (S-SPECT) camera. Results DNM sensitivity was more than threefold higher while SR was notably better. Significantly, SR was the same for 99mTc and 201Tl. ER was improved on DNM and allowed good separation of 99mTc and 123I spectral peaks. Count rate remained linear on DNM up to 612 kcps, while S-SPECT showed severe dead time limitations. Phantom studies revealed comparable uniformity and defect contrast, notwithstanding significantly shorter acquisition time for the DNM. First patient images, including dynamic SPECT, are also presented. Conclusion DNM is raising the bar for expedition and upgrade of practice. It features high sensitivity as well as improved SR, temporal resolution and ER. It enables reduction of acquisition time and fast protocols. Importantly, it is potentially capable of dynamic 3-D acquisition. The new technology is potentially upgradeable and may become a milestone in the evolution of nuclear cardiology as it assumes its key role in molecular imaging of the heart.Keywords
This publication has 78 references indexed in Scilit:
- Novel solid-state-detector dedicated cardiac camera for fast myocardial perfusion imaging: multicenter comparison with standard dual detector camerasJournal of Nuclear Cardiology, 2009
- Breaking America's Dependence on Imported MolybdenumJACC: Cardiovascular Imaging, 2009
- Radionuclide ImagingJACC, 2008
- Technological Development and Advances in Single-Photon Emission Computed Tomography/Computed TomographySeminars in Nuclear Medicine, 2008
- Clinical Myocardial Perfusion PET/CTJournal of Nuclear Medicine, 2007
- Comparison Between Angiography and Fractional Flow Reserve Versus Single-Photon Emission Computed Tomographic Myocardial Perfusion Imaging for Determining Lesion Significance in Patients With Multivessel Coronary DiseaseThe American Journal of Cardiology, 2007
- A multipinhole small animal SPECT system with submillimeter spatial resolutionMedical Physics, 2006
- Myocardial perfusion scintigraphy and cost effectiveness of diagnosis and management of coronary heart diseaseHeart, 2004
- Assessment of Right Ventricular Morphology and FunctionSeminars in Respiratory and Critical Care Medicine, 2003
- Compartmental Modeling of Technetium-99m–Labeled Teboroxime with Dynamic Single-Photon Emission Computed TomographyInvestigative Radiology, 2001