Oxygen electrode response lag induced by liquid film resistance against oxygen transfer
Authors:
- 1 November 1976
- journal article
- research article
- Published by Wiley in Biotechnology and Bioengineering
- Vol. 18 (11) , 1537-1555
Abstract
The response of a polarographic oxygen electrode to a step change and to an exponential change in bulk oxygen concentration was studied theoretically and experimentally for the case where there is a significant liquid film resistance at the outer side of the membrane‐covered electrode. The probe response has been described by a mathematical model which takes into account the oxygen mass transfer through the liquid film and through the membrane. It is shown that substantial error could be introduced by neglecting the liquid film resistance. A test procedure is suggested to evaluate the error significance.Keywords
This publication has 7 references indexed in Scilit:
- Oxygen absorption into glucose solutionChemical Engineering Science, 1972
- Determination of aeration capacity of mechanically. Agitated vessels by fast response oxygen probeBiotechnology and Bioengineering, 1972
- Oxygen mass transfer and oxygen respiration rate measurements utilizing fast response oxygen electrodesBiotechnology and Bioengineering, 1971
- Polarographic oxygen analyzer response: The effect of instrument lag in the non-steady state reaeration testWater Research, 1970
- Dynamic measurement of the volumetric oxygen transfer coefficient in fermentation systemsBiotechnology and Bioengineering, 1967
- Design and application of electrodes for the determination of dissolved oxygenFolia Microbiologica, 1965
- Continuous Recording of Blood Oxygen Tensions by PolarographyJournal of Applied Physiology, 1953