Bismuth Shielding, Organ-based Tube Current Modulation, and Global Reduction of Tube Current for Dose Reduction to the Eye at Head CT
Authors:
- Xinhui DuanXDXinhui Duan
- Jodie A. ChristnerJCJodie A. Christner
- Shuai LengSLShuai Leng
- Katharine L. GrantKGKatharine L. Grant
- Cynthia H. McColloughCMCynthia H. McCollough
- 1 January 2012
- journal article
- research article
- Published by Radiological Society of North America (RSNA) in Radiology
- Vol. 262 (1) , 191-198
Abstract
Organ-based tube current modulation provided similar dose reduction to the eye as bismuth shielding but better image quality; a global reduction of tube current by approximately 30% achieves the same dose reduction to the eye as bismuth shielding but also reduces dose to the brain and maintains CT number accuracy. PurposeTo compare the dose and image quality of three methods for reducing the radiation dose to the eye at head computed tomography (CT): bismuth shielding, organ-based tube current modulation (TCM), and global reduction of the tube current.Materials and MethodsAn anthropomorphic head phantom was scanned under six conditions: (a) without any dose reduction techniques (reference scanning); (b) with one bismuth eye shield; (c) with organ-based TCM; (d) with reduced tube current to yield the same dose reduction as one bismuth shield; (e) with two layers of bismuth shields; and (f) with organ-based TCM and one bismuth shield. Dose to the eye, image noise, and CT numbers in the brain region were measured and compared. The effect of increasing distance between the bismuth shield and eye lens was also investigated.ResultsRelative to the reference scan, the dose to the eye was reduced by 26.4% with one bismuth shield, 30.4% with organ-based TCM, and 30.2% with a global reduction in tube current. A combination of...Keywords
This publication has 26 references indexed in Scilit:
- Lens Exposure during Brain Scans Using Multidetector Row CT Scanners: Methods for Estimation of Lens DoseAmerican Journal of Neuroradiology, 2009
- Thyroid Shields versus z-Axis Automatic Tube Current Modulation for Dose Reduction at Neck CTRadiology, 2008
- Comparison of Eye Lens Dose on Neuroimaging Protocols between 16- and 64-Section Multidetector CT: Achieving the Lowest Possible DoseAmerican Journal of Neuroradiology, 2008
- Risk of Cataract after Exposure to Low Doses of Ionizing Radiation: A 20-Year Prospective Cohort Study among US Radiologic TechnologistsAmerican Journal of Epidemiology, 2008
- MOSFET Dosimetry for Radiation Dose Assessment of Bismuth Shielding of the Eye in ChildrenAmerican Journal of Roentgenology, 2007
- AbstractAnnals of the ICRP, 2007
- Cataracts among Chernobyl Clean-up Workers: Implications Regarding Permissible Eye ExposuresRadiation Research, 2007
- Orbital, thyroid, and breast superficial radiation shielding for patients undergoing diagnostic CTSeminars in Ultrasound, CT and MRI, 2002
- A clinical study of radiation cataract formation in adult life following gamma irradiation of the lens in early childhoodBritish Journal of Ophthalmology, 1997
- IntroductionAnnals of the ICRP, 1991