Translation of First North American 50 and 70 cc Total Artificial Heart Virtual and Clinical Implantations: Utility of 3D Computed Tomography to Test Fit Devices
Authors:
- 10 November 2016
- journal article
- case report
- Published by Wiley in Artificial Organs
- Vol. 41 (8) , 727-734
Abstract
Since the creation of SynCardia's 50 cc Total Artificial Hearts (TAHs), patients with irreversible biventricular failure now have two sizing options. Herein, a case series of three patients who have undergone successful 50 and 70 cc TAH implantation with complete closure of the chest cavity utilizing preoperative “virtual implantation” of different sized devices for surgical planning are presented. Computed tomography (CT) images were used for preoperative planning prior to TAH implantation. Three-dimensional (3D) reconstructions of preoperative chest CT images were generated and both 50 and 70 cc TAHs were virtually implanted into patients' thoracic cavities. During the simulation, the TAHs were projected over the native hearts in a similar position to the actual implantation, and the relationship between the devices and the atria, ventricles, chest wall, and diaphragm were assessed. The 3D reconstructed images and virtual modeling were used to simulate and determine for each patient if the 50 or 70 cc TAH would have a higher likelihood of successful implantation without complications. Subsequently, all three patients received clinical implants of the properly sized TAH based on virtual modeling, and their chest cavities were fully closed. This virtual implantation increases our confidence that the selected TAH will better fit within the thoracic cavity allowing for improved surgical outcome. Clinical implantation of the TAHs showed that our virtual modeling was an effective method for determining the correct fit and sizing of 50 and 70 cc TAHs.Keywords
This publication has 13 references indexed in Scilit:
- First North American 50 cc Total Artificial Heart Experience: Conversion from a 70 cc Total Artificial HeartASAIO Journal, 2016
- Virtual implantation of the 50cc SynCardia total artificial heartThe Journal of Heart and Lung Transplantation, 2016
- 2013 ACCF/AHA Guideline for the Management of Heart Failure: Executive SummaryCirculation, 2013
- Survival after biventricular mechanical circulatory support: Does the type of device matter?The Journal of Heart and Lung Transplantation, 2012
- 377 Survival to Transplant in Patients Undergoing Mechanical Circulatory Support as Bridge: Retrospective Analysis of LVAD, BiVAD and TAH StrategyThe Journal of Heart and Lung Transplantation, 2012
- The Fourth INTERMACS Annual Report: 4,000 implants and countingThe Journal of Heart and Lung Transplantation, 2012
- Current Trends in Implantable Left Ventricular Assist DevicesCardiology Research and Practice, 2011
- Mechanical Device–Based Methods of Managing and Treating Heart FailureCirculation, 2005
- Cardiac Replacement with a Total Artificial Heart as a Bridge to TransplantationNew England Journal of Medicine, 2004
- ACC/AHA Guidelines for the Evaluation and Management of Chronic Heart Failure in the Adult: Executive Summary A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines (Committee to Revise the 1995 Guidelines for the Evaluation and Management of Heart Failure)Circulation, 2001