Development and performance evaluation of an experimental fine pitch detector multislice CT scanner

Authors:

Abstract
The authors have developed an experimental fine pitch detector multislice CT scanner with an ultrasmall focal spot x‐ray tube and a high‐density matrix detector through current CT technology. The latitudinal size of the x‐ray tube focal spot was . The detector dimension was 1824 channels rows (longitudinal direction) at row width of , in which a thinner reflected separator surrounds each detector cell coupled with a large active area photodiode. They were mounted on a commercial 64‐slice CT scanner gantry while the scan field of view and gantry rotation speed can be maintained. The experimental CT scanner demonstrated the spatial resolution of with the acrylic slit phantom and in‐plane 50%‐MTF and 10%‐MTF . In the longitudinal direction, it demonstrated the spatial resolution of with the high‐resolution insert of the CATPHAN phantom and as the full width at half maximum of the slice sensitivity profile. In low‐contrast detectability, at 0.3% was visualized at the of . Two types of diameter vessel phantoms with in‐stent stenosis at 25%, 50%, and 75% stair steps were scanned, and the reconstructed images can clearly resolve the stenosis at each case. The experimental CT scanner provides high‐resolution imaging while maintaining low‐contrast detectability, demonstrating the potentiality for clinical applications demanding high spatial resolution, such as imaging of inner ear, lung, and bone, or low‐contrast detectability, such as imaging of coronary artery.