High-power short-duration catheter ablation of atrial fibrillation: is it really a new era? Comparison between new and old radiofrequency contact force–sensing catheters
- 16 August 2023
- journal article
- research article
- Published by Springer Nature in Journal of Interventional Cardiac Electrophysiology
- Vol. 67 (3) , 549-557
Abstract
Background The clinical performance of high-power, short-duration (HPSD) pulmonary vein isolation (PVI) with the novel flexible tip TactiFlex™ (TFSE) catheter, as compared to standard-power, long-duration (SPLD) PVI using the TactiCath™ (TCSE) catheter among patients undergoing catheter ablation (CA) of atrial fibrillation (AF) is currently unknown. Methods We conducted a prospective, observational, single-centre study including 40 consecutive patients undergoing PVI for paroxysmal/persistent AF, using HPSD ablation with the novel TFSE catheter (HPSD/TFSE group). Based on propensity score-matching, forty patients undergoing SPLD PVI with the TCSE catheter were identified (SPLD/TCSE group). In the HPSD/TFSE group, RF lesions were performed by delivering 40–50 W for 10–20 s, while in the SPLD/TCSE group, RF power was 30–35 W, targeting a lesion size index (LSI) of 4.0–5.5. The co-primary study outcomes were time required to complete PVI and first pass isolation (FPI). Results PVI was achieved in 100% of patients in both groups, and no major adverse events were observed. Remarkably, PVI time was shorter in the HPSD/TFSE, compared to the SPLD/TCSE group(9 [7–9] min vs. 50 [37–54] min; p < 0.001), while FPI rate was non-significantly higher in the former group(91% [146/160] vs 83% [134/160]; p = 0.063). Shorter procedural (108 [91–120] min vs. 173 [139–187] min, p < 0.001), total RF (9 [7–11] min vs. 43 [32–53] min, p < 0.001), fluoroscopy times(15 [10–19] min vs. 18 [13–26] min, p = 0.014), and lower DAP (1461 [860–2181] vs. 7200 [3400–20,800], p < 0.001) were recorded in the HPSD/TFSE group. A higher average impedance drop was obtained with HPSD/TFSE CA(17[17–18]Ω vs. 16 [15–17] Ω, p < 0.001). Conclusions In our initial clinical experience, HPSD PVI with the TFSE catheter proved faster than SPLD PVI with the TCSE catheter, at least equally effective in terms of FPI, and it was associated with greater impedance drop. Graphical abstract Main study findings. Abbreviations: DAP, dose area product; HPSD–TFSE, high-power short-duration with the TactiFlex™, Sensor Enabled™ catheter; PVI, pulmonary vein isolation; RF, radiofrequency energy; SPLD–TCSE, standard-power long-duration ablation with the TactiCath, Sensor Enabled™ catheterKeywords
This publication has 30 references indexed in Scilit:
- First clinical use of novel ablation catheter incorporating local impedance dataJournal of Cardiovascular Electrophysiology, 2018
- High-Power and Short-Duration Ablation for Pulmonary Vein IsolationJACC: Clinical Electrophysiology, 2018
- Atrial fibrillation ablation using very short duration 50 W ablations and contact force sensing cathetersJournal of Interventional Cardiac Electrophysiology, 2018
- 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillationEP Europace, 2017
- In Vitro Validation of the Lesion Size Index to Predict Lesion Width and Depth After Irrigated Radiofrequency Ablation in a Porcine ModelJACC: Clinical Electrophysiology, 2017
- Long-term efficacy of catheter ablation as first-line therapy for paroxysmal atrial fibrillation: 5-year outcome in a randomised clinical trialHeart, 2016
- Ablation Versus Amiodarone for Treatment of Persistent Atrial Fibrillation in Patients With Congestive Heart Failure and an Implanted DeviceCirculation, 2016
- Initial impedance decrease as an indicator of good catheter contact: Insights from radiofrequency ablation with force sensing cathetersHeart Rhythm, 2014
- A novel radiofrequency ablation catheter using contact force sensing: Toccata studyHeart Rhythm, 2012
- Pulmonary Vein Antral Isolation Using an Open Irrigation Ablation Catheter for the Treatment of Atrial FibrillationJACC, 2007