Multislice imaging of quantitative cerebral perfusion with pulsed arterial spin labeling
Authors:
- 1 May 1998
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 39 (5) , 825-832
Abstract
A method is presented for multislice measurements of quantitative cerebral perfusion based on magnetic labeling of arterial spins. The method combines a pulsed arterial inversion, known as the FAIR (Flow‐sensitive Alternating Inversion Recovery) experiment, with a fast spiral scan image acquisition. The short duration (22 ms) of the spiral data collection allows simultaneous measurement of up to 10 slices per labeling period, thus dramatically increasing efficiency compared to current single slice acquisition protocols. Investigation of labeling efficiency, suppression of unwanted signals from stationary as well as intraarterial spins, and the FAIR signal change as a function of inversion delay are presented. The assessment of quantitative cerebral blood flow (CBF) with the new technique is demonstrated and shown to require measurement of arterial transit time as well as suppression of intraarterial spin signals. CBF values measured on normal volunteers are consistent with results obtained from H2O15 positron emission tomography (PET) studies and other radioactive tracer approaches. In addition, the new method allows detection of activation‐related perfusion changes in a finger‐tapping experiment, with locations of activation corresponding well to those observed with blood oxygen level dependent (BOLD) fMRI.Keywords
This publication has 29 references indexed in Scilit:
- Quantitation of Regional Cerebral Blood Flow Increases during Motor Activation: A Steady-State Arterial Spin Tagging StudyNeuroImage, 1997
- T1 and T2 of ferritin solutions: Effect of loading factorMagnetic Resonance in Medicine, 1996
- Analysis of fMRI Time-Series Revisited—AgainNeuroImage, 1995
- Motion Artifacts in fMRI: Comparison of 2DFT with PR and Spiral Scan MethodsMagnetic Resonance in Medicine, 1995
- 19F Magnetic Resonance Imaging of Cerebral Blood Flow with 0.4-cc ResolutionJournal of Cerebral Blood Flow & Metabolism, 1994
- Inflow versus deoxyhemoglobin effects in bold functional MRI using gradient echoes at 1.5 TNMR in Biomedicine, 1994
- Tissue specific perfusion imaging using arterial spin labelingNMR in Biomedicine, 1994
- Perfusion imagingMagnetic Resonance in Medicine, 1992
- Functional Mapping of the Human Visual Cortex by Magnetic Resonance ImagingScience, 1991
- Concurrent measurements of cerebral blood flow, sodium, lactate, and high‐energy phosphate metabolism using 19F, 23Na, 1H, and 31P nuclear magnetic resonance spectroscopyMagnetic Resonance in Medicine, 1988