Diagnostic Accuracy and Safety of Confirm Rx™ Insertable Cardiac Monitor in Pediatric Patients
- 9 October 2020
- journal article
- research article
- Published by Springer Nature in Pediatric Cardiology
- Vol. 42 (1) , 142-147
Abstract
Insertable cardiac monitors (ICM) are subcutaneously implanted devices that monitor a patient’s heart rate and rhythm (Rossano in Pediatrics 112(3):e228, 2003). The diagnostic accuracy and safety of the Confirm RxTM (Abbott, Minneapolis, MN) ICM in pediatric patients is unknown. This is a single center, retrospective, IRB-approved review of patients ≤ 21 years implanted with Confirm RxTM ICMs from 2017 to 2020. Data collected included demographics, indications, presence of P-wave and R-wave amplitude at implantation and follow-up, number/appropriateness of transmissions pre and post implementation of SharpSenseTM technology, reprogramming to improve accuracy, time from implantation to arrhythmia detection, and complications. There were 29 patients (median age: 8 years, 59% females). P-waves were identified in all patients and average R-wave amplitude was 0.85 mV (0.26–1.03 mV). There was no significant difference in R-wave amplitude based on size (BSA ≥ 1.5 m2: 0.76 mV, < 1.5 m2: 0.91 mV) or congenital heart disease (+CHD: 0.86 mV, −CHD: 0.85 mV). Arrhythmias identified were the following: wide complex tachycardia (1), supraventricular tachycardia (4), bradycardia/sinus pause (3), and premature ventricular contraction (1). SharpSenseTM implementation decreased the false-positive rate in device-initiated transmissions (55.4% to 0%, p < 0.00001). Average time from implantation to arrhythmia detection was 2.63 months (range: 0–8.8). A single complication of cellulitis occurred. Confirm RxTM is appropriate for implant in pediatric patients regardless of age, BSA, or CHD. Implementation of SharpSense™ technology dramatically decreased the false-positive rate. Follow-up studies could utilize additional monitoring devices to provide analysis on potential events that the Confirm RxTM ICM missed.Keywords
This publication has 10 references indexed in Scilit:
- The Real-World Utility of the LINQ Implantable Loop Recorder in Pediatric and Adult Congenital Heart PatientsJACC: Clinical Electrophysiology, 2019
- Insertable cardiac monitors: current indications and devicesExpert Review of Medical Devices, 2018
- Implantable loop recorder in clinical practiceJournal of Arrhythmia, 2018
- Current and Future Use of Insertable Cardiac MonitorsJACC: Clinical Electrophysiology, 2018
- Diagnostic and therapeutic value of implantable loop recorder: A tertiary care center experiencePacing and Clinical Electrophysiology, 2018
- Implantable Loop Recorder in Inherited Arrhythmia DiseasesJACC: Clinical Electrophysiology, 2018
- Neonatal and Pediatric ArrhythmiasCardiac Electrophysiology Clinics, 2018
- Utility and limitations of long-term monitoring of atrial fibrillation using an implantable loop recorderHeart Rhythm, 2018
- P- and R-Wave Amplitude Sensed By Reveal LINQ™ Loop Recorder in Pediatric PatientsJournal of Innovations in Cardiac Rhythm Management, 2017