Revision of Broken Knee Megaprostheses: New Solution to Old Problems
Authors:
- 1 November 2010
- journal article
- research article
- Published by Wolters Kluwer Health in Clinical Orthopaedics and Related Research
- Vol. 468 (11) , 2904-2913
Abstract
Eakage. We also report our experience with revision surgery. Methods We identified 28 breakages in 266 megaprosthetic knee arthroplasties performed between January2000 and December 2006. Twenty-six breakages were revised to another prosthesis. The complications were studied and the function was evaluated. Prostheses were studied for failure by the computer-aided design program SolidWorks® and Hyperworks® for finite element analysis (FEA). Design improvements were performed based on these results. Results In 21 cases, the failure occurred at the stem-collar junction, the point of maximum stress predicted by FEA. Stainless steel implants were prone to failure. There was one early and one late infection. Three patients died of metastatic disease. The most difficult surgical step involved the removal of the well-cemented broken stem from the intramedullary canal. Musculoskeletal Tumor Society scores varied from 27 to 29 after revision. FEA revealed stress could be reduced by filleting the stem-collar junction and by two-piece stems. Conclusions Revisions of broken total knee megaprostheses, though technically difficult, have allowed patients reasonable function. We recommend design analysis for custom prostheses to point to areas of weakness. Breakages can be reduced by using titanium stems and filleting the junction or by having two-piece inserted stems. Incorporating these changes has reduced the failures in our experience. Level of Evidence Level IV, therapeutic study. See Guidelines for Authors for a complete description of levels of evidence....Keywords
This publication has 24 references indexed in Scilit:
- Endoprosthetic replacement for primary tumours around the kneeThe Journal of Bone and Joint Surgery. British volume, 2008
- Limb Salvage Surgery for OsteosarcomaClinical Orthopaedics and Related Research, 2007
- Endoprosthetic replacement of the distal femur for bone tumoursThe Journal of Bone and Joint Surgery. British volume, 2007
- Endoprosthetic Reconstruction in 250 Patients with SarcomaClinical Orthopaedics and Related Research, 2006
- Aseptic Loosening is Uncommon with Uncemented Proximal Tibia Tumor ProsthesesClinical Orthopaedics and Related Research, 2006
- Cemented Rotating Hinge Endoprosthesis for Limb Salvage of Distal Femur TumorsClinical Orthopaedics and Related Research, 2006
- Survivorship and clinical outcome of modular endoprosthetic reconstruction for neoplastic disease of the lower limbThe Journal of Bone and Joint Surgery. British volume, 2006
- Distal Femur Replacement Is Useful in Complex Total Knee Arthroplasty RevisionsClinical Orthopaedics and Related Research, 2006
- Revision of the Kotz type of tumour endoprosthesis for the lower limbThe Journal of Bone and Joint Surgery. British volume, 2002
- Megaprostheses after resection of distal femoral tumors: A rotating hinge design in 30 patients followed for 2-7 yearsActa Orthopaedica, 1996