Fabrication of bridge-type microcalorimeter arrays with Ti-Au transition-edge-sensors
Authors:
- 8 March 2002
- conference paper
- Published by AIP Publishing in AIP Conference Proceedings
- Vol. 605 (1) , 207-210
Abstract
The present stage of cryogenic detector development is focused on fabrication of arrays. Arrays of superconducting-transition-edge-sensor (TES) microcalorimeters have generally been fabricated on membranes. In conventional membrane structures fabricated by backside etching, it is difficult to realize arrays having both mechanical toughness and a large filling factor. It is possible to overcome these difficulties by using a bridge-type membrane structure, in which membranes are raised above the Si substrate with a gap of 10–50 μm. In order to carry out etching and other processes, we have fabricated a small scale array with a bridge-type membrane structure. A preliminary version of a TES microcalorimeter array with a pixel structure had dimensions of Each pixel consisted of a Ti/Au bilayer with a size of on a 1 μm-thick membrane of Although the current array exhibited a small filling factor of about 0.1, it is possible to obtain a high filling factor by fabricating mushroom-shaped absorbers.