Abstract
TLD-300 (CaF2:Tm) dosimeters were used to measure the absorbed dose in an experimental assembly simulating the shield and the superconducting coils of a fusion reactor irradiated by 14-MeV neutrons. The shield was formed by plates of Type 316 stainless steel and by a water-equivalent material (Perspex), while a second block made of Type 316 stainless steel and copper plates simulated the superconducting coils of the tokamak. Since the TLD-300 shows two main peaks, one of which is more sensitive to neutrons, the neutron and gamma doses were separated using the two-peak method. The resulting absorbed neutron dose was 30% of the total in positions close to the neutron source, while its contribution was negligible (< 7%) in the superconducting coils. The total dose level to be studied ranged from a few tens of micrograys to 10 Gy. Because the latter value was expected to be out of the linear response range for the TLD-300, the supralinear effects for the TLD-300 were studied as well as its sensitivity...