A simple model of the effects of atmospheric absorption and emission on radiation thermometer readings is described, and methods of correcting for these effects are given. These methods are applied to measurements made in a large reformer furnace using a commercially available radiation thermometer operating near 1 μm, whose spectral responsivity partially overlaps some of the water vapor absorption bands. An effective absorption coefficient for the thermometer and the flue gas temperature are calculated and used to correct the thermometer readings. For relatively long measurement path lengths the corrections are almost as significant as reflection errors, and the uncertainties in the corrections increase very rapidly with path length.