A simplified biventricular defibrillator with fixed long detection intervals reduces implantable cardioverter defibrillator (ICD) interventions and heart failure hospitalizations in patients with non-ischaemic cardiomyopathy implanted for primary prevention: the RELEVANT [Role of long dEtection window programming in patients with LEft VentriculAr dysfunction, Non-ischemic eTiology in primary prevention treated with a biventricular ICD] study
Open Access
- 29 June 2009
- journal article
- research article
- Published by Oxford University Press (OUP) in European Heart Journal
- Vol. 30 (22) , 2758-2767
Abstract
To investigate the efficacy and safety of a cardiac resynchronization therapy with cardioverter–defibrillator (CRT-D) device with simplified ventricular tachycardia management in patients with non-ischaemic heart failure (HF) and primary prevention implantable cardioverter defibrillator (ICD) indication. Prospective, controlled, parallel, multicentre, non-randomized study enrolling 324 primary prevention non-ischaemic HF patients implanted with CRT-D devices from 2004 to 2007: Protect group, 164 patients implanted with a Medtronic Insync III Protect device and Control group, 160 patients utilizing other Medtronic CRT-D devices. Efficacy was assessed by computing appropriate and inappropriate detections and therapies during follow-up; safety compared hospitalizations and syncopal events between groups. Ninety per cent of both ventricular and supraventricular tachyarrhythmias terminated within the 13–29 beat detection interval with the Protect algorithm. The Protect group showed a significantly better event-free survival to first delivered therapy for total (P = 0.0001), appropriately treated (P = 0.002), and inappropriately treated episodes (P = 0.017). The total number of delivered shocks was significantly lower in the Protect group (22 vs. 59, P < 0.0001). In the Protect group, a significantly reduced HF hospitalization (hazard ratio 0.38, 95% CI 0.15–0.98, P = 0.044) was observed without any increase of syncope or death. A simplified CRT-D device with fixed long detection reduced overall ICD therapy burden and HF hospitalizations without entailing any additional adverse events in primary prevention non-ischaemic HF patients.Keywords
This publication has 28 references indexed in Scilit:
- Prognostic Importance of Defibrillator Shocks in Patients with Heart FailureNew England Journal of Medicine, 2008
- Use of Cardiac Resynchronization Therapy in Patients Hospitalized With Heart FailureCirculation, 2008
- Guidelines for cardiac pacing and cardiac resynchronization therapy: The Task Force for Cardiac Pacing and Cardiac Resynchronization Therapy of the European Society of Cardiology. Developed in Collaboration with the European Heart Rhythm AssociationEuropean Heart Journal, 2007
- Characteristics of ventricular tachyarrhythmias occurring in ischemic versus nonischemic patients implanted with a biventricular cardioverter-defibrillator for primary or secondary prevention of sudden deathAmerican Heart Journal, 2006
- A Comparison of Empiric to Physician-Tailored Programming of Implantable Cardioverter-DefibrillatorsJACC, 2006
- Differences in Tachyarrhythmia Detection and Implantable Cardioverter Defibrillator Therapy by Primary or Secondary Prevention Indication in Cardiac Resynchronization Therapy PatientsJournal of Cardiovascular Electrophysiology, 2004
- Classification of death in antiarrhythmia trialsJACC, 1996
- Implantable Defibrillation and Thromboembolic EventsPacing and Clinical Electrophysiology, 1995
- Hemodynamic deterioration during ICD implant: Predictors of high-risk patientsAmerican Heart Journal, 1994
- Clinical Performance of the Implantable Cardioverter Defibrillator: Electrocardiographic Documentation of 101 Spontaneous DischargesPacing and Clinical Electrophysiology, 1991