Design and Characterization of a Multisource Hand-Held Tandem Mass Spectrometer
- 29 August 2008
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 80 (19) , 7198-7205
Abstract
A wireless-controlled miniature rectilinear ion trap mass spectrometer system, total weight with batteries 5.0 kg, consuming less than 35 W of power, and having dimensions of 22 cm in length by 12 cm in width by 18 cm in height, is characterized. The design and construction of the mass spectrometer including mass analyzer, vacuum system, electronics system, and data acquisition and processing systems, is detailed. The mass spectrometer is compatible with various types of ionization sources including a glow discharge electron impact ionization source used in the internal ionization mode, and various atmospheric pressure ionization sources, including electrospray ionization, atmospheric pressure chemical ionization, and desorption electrospray ionization, which are employed for external, atmospheric pressure ionization. These external sources are coupled to the miniature mass spectrometer via a capillary interface that is operated in a discontinuous fashion (discontinuous atmospheric pressure interface) to maximize ion transport. The performance of the mass spectrometer for large and small molecules is characterized. Limits of detection in the parts-per-billion range were obtained for selected compounds examined using both the internal ionization and external ionization modes. Tandem mass spectrometry and fast in situ analysis capabilities are also demonstrated using a variety of compounds and ionization sources. Protein molecules are analyzed as the multiply protonated molecules with mass/charge ratios up to 1500 Da/charge.Keywords
This publication has 29 references indexed in Scilit:
- Halo Ion Trap Mass SpectrometerAnalytical Chemistry, 2007
- Handheld Rectilinear Ion Trap Mass SpectrometerAnalytical Chemistry, 2006
- Miniature toroidal radio frequency ion trap mass analyzerJournal of the American Society for Mass Spectrometry, 2006
- Fabrication and testing of a miniature cylindrical ion trap mass spectrometer constructed from low temperature co-fired ceramicsInternational Journal of Mass Spectrometry, 2006
- Rectilinear Ion Trap: Concepts, Calculations, and Analytical Performance of a New Mass AnalyzerAnalytical Chemistry, 2004
- Miniature Cylindrical Ion Trap Mass SpectrometerAnalytical Chemistry, 2002
- A two-dimensional quadrupole ion trap mass spectrometerJournal of the American Society for Mass Spectrometry, 2002
- Miniature mass analyzersJournal of Mass Spectrometry, 2000
- A Miniature Cylindrical Quadrupole Ion Trap: Simulation and ExperimentAnalytical Chemistry, 1998
- Miniature, high-resolution, quadrupole mass-spectrometer arrayReview of Scientific Instruments, 1997