Measurement of helium-like and hydrogen-like argon spectra using double-crystal X-ray spectrometers on EAST
- 18 August 2016
- journal article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 87 (11) , 11E326
Abstract
A two-crystal assembly was deployed on the tangential X-ray crystal spectrometer to measure both helium-like and hydrogen-like spectra on EAST. High-quality helium-like and hydrogen-like spectra were observed simultaneously for the first time on one detector for a wide range of plasma parameters. Profiles of line-integrated core ion temperatures inferred from two spectra were consistent. Since tungsten was adopted as the upper divertor material, one tungsten line (W XLIV at 4.017 Å) on the short-wavelength side of the Lyman-α line (Lα1) was identified for typical USN discharges, which was diffracted by a He-like crystal (2d = 4.913 Å). Another possible Fe XXV line (1.85 Å) was observed to be located on the long-wavelength side of resonance line (w), which was diffracted from a H-like crystal (2d = 4.5622 Å) on the second order. Be-like argon lines were also observable that fill the detector space between the He-like and H-like spectra.Keywords
This publication has 5 references indexed in Scilit:
- Upgrades of imaging x-ray crystal spectrometers for high-resolution and high-temperature plasma diagnostics on EASTReview of Scientific Instruments, 2014
- Space-resolved extreme ultraviolet spectrometer system for impurity behavior research on experimental advanced superconducting TokamakNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2013
- X-ray imaging crystal spectroscopy for use in plasma transport researchReview of Scientific Instruments, 2012
- X-ray observations of 2l-nl’ transitions in–from tokamak plasmasPhysical Review A, 1995