The Correlation of Pupil Size Measured by Colvard Pupillometer and Orbscan II
Authors:
- 1 October 2007
- journal article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 23 (10) , 789-795
Abstract
Purpose: To assess the correlation of pupil sizes measured under different illuminations by Colvard pupillometer and the Orbscan II system and to determine the factors that may influence pupil sizes. Methods: The study comprised 162 eyes of 162 patients with a mean age of 34.6±10.6 years (range: 16 to 65 years). The photopic pupil sizes were measured with Orbscan II (363 lux). The high mesopic and low mesopic pupil sizes were measured by Colvard pupillometer under 60.5 and 0.15 lux of illumination, respectively. Spherical equivalent of the manifest refraction was used to evaluate its correlation with the pupil size. Results: The mean sizes of low mesopic and high mesopic pupil measured by Colvard pupillometer were 6.08±0.68 mm and 4.80±0.79 mm, respectively. The mean pupil size measured by Orbscan II was 3.87±0.61 mm. The pupil sizes were negatively correlated with age, but not correlated with refractive error when using multiple linear regression. Low mesopic pupil sizes were only moderately correlated with pupil sizes measured by Orbscan II (low mesopic pupil size [mm] = 0.755 × Orbscan II pupil size [mm] + 3.13, r = 0.67). If >4.5-mm pupil size measured by Orbscan II is used as the cut-off value for detecting low mesopic pupil sizes >6.5 mm, there will be a sensitivity of 48.15% and a specificity of 93.97%. Conclusions: Pupil sizes measured with the Orbscan II were only moderately correlated with low mesopic pupil sizes measured with Colvard pupillometer. The Orbscan II is not a preferred tool for estimating low mesopic pupil sizes. [ J Refract Surg . 2007;23:789–795.]Keywords
This publication has 20 references indexed in Scilit:
- Risk factors for night vision complaints after LASIK for myopiaOphthalmology, 2004
- Pupil size and quality of vision after LASIKOphthalmology, 2003
- Quality of vision after laser in situ keratomileusisJournal of Cataract & Refractive Surgery, 2003
- Comparison of scotopic pupil measurement with slitlamp-based cobalt blue light and infrared video-based systemJournal of Cataract & Refractive Surgery, 2002
- Use of a digital infrared pupillometer to assess patient suitability for refractive surgeryJournal of Cataract & Refractive Surgery, 2002
- LettersJournal of Cataract & Refractive Surgery, 1998
- LettersJournal of Cataract & Refractive Surgery, 1998
- Inter-observer variation in the evaluation of neurological signs: Patient-related factorsJournal of Neurology, 1994
- The effect of pupil centration and diameter on ocular performanceVision Research, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986