Components of respiratory resistance monitored in mechanically ventilated patients

Authors:
  • FPF. Peták
    F. Peták
  • TAT. Asztalos
    T. Asztalos
  • ZDZ.I. Deák
    Z.I. Deák
  • GBG. Bogáts
    G. Bogáts
  • ZHZ. Hantos
    Z. Hantos

Abstract
The interrupter technique is commonly adopted to monitor respiratory resistance (Rrs,int) during mechanical ventilation; however,Rrs,intis often interpreted as an index of airway resistance (Raw).This study compared the values ofRrs,intprovided by a Siemens 940 Lung Mechanics Monitor with total respiratory impedance (Zrs) parameters in 39 patients with normal spirometric parameters, who were undergoing elective coronary bypass surgery.Zrswas determined at the airway opening with pseudorandom oscillations of 0.2–6 Hz at end inspiration.Rawand tissue resistance (Rti) were derived from theZrsdata by model fitting;Rtiand total resistance (Rrs,osc=Raw+Rti) were calculated at the actual respirator frequencies. Lower airway resistance (Rawl) was estimated by measuring tracheal pressure.Although good agreement was obtained betweenRrs,oscandRrs,int, with a ratio of 1.07±0.19 (mean±sd), they correlated poorly (r2=0.36).Rtiand the equipment component ofRawaccounted for most ofRrs,osc(39.8±11.9 and 43.0±6.9%, respectively), whereas only a small portion belonged toRawl(17.2±6.3%).It is concluded that respiratory resistance may become very insensitive to changes in lower airway resistance and therefore, inappropriate for following alterations in airway tone during mechanical ventilation, especially in patients with relatively normal respiratory mechanics, where the tissue and equipment resistances represent the vast majority of the total resistance.