Feasibility of a GATE Monte Carlo platform in a clinical pretreatment QA system for VMAT treatment plans using TrueBeam with an HD120 multileaf collimator
Authors:
Open Access
- 23 September 2019
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 20 (10) , 101-110
Abstract
Purpose: To evaluate the quality of patient‐specific complicated treatment plans, including commercialized treatment planning systems (TPS) and commissioned beam data, we developed a process of quality assurance (QA) using a Monte Carlo (MC) platform. Specifically, we constructed an interface system that automatically converts treatment plan and dose matrix data in digital imaging and communications in medicine to an MC dose‐calculation engine. The clinical feasibility of the system was evaluated.Materials and Methods: A dose‐calculation engine based on GATE v8.1 was embedded in our QA system and in a parallel computing system to significantly reduce the computation time. The QA system automatically converts parameters in volumetric‐modulated arc therapy (VMAT) plans to files for dose calculation using GATE. The system then calculates dose maps. Energies of 6 MV, 10 MV, 6 MV flattening filter free (FFF), and 10 MV FFF from a TrueBeam with HD120 were modeled and commissioned. To evaluate the beam models, percentage depth dose (PDD) values, MC calculation profiles, and measured beam data were compared at various depths (Dmax, 5 cm, 10 cm, and 20 cm), field sizes, and energies. To evaluate the feasibility of the QA system for clinical use, doses measured for clinical VMAT plans using films were compared to dose maps calculated using our MC‐based QA system.Results: A LINAC QA system was analyzed by PDD and profile according to the secondary collimator and multileaf collimator (MLC). Values for MC calculations and TPS beam data obtained using CC13 ion chamber (IBA Dosimetry, Germany) were consistent within 1.0%. Clinical validation using a gamma index was performed for VMAT treatment plans using a solid water phantom and arbitrary patient data. The gamma evaluation results (with criteria of 3%/3 mm) were 98.1%, 99.1%, 99.2%, and 97.1% for energies of 6 MV, 10 MV, 6 MV FFF, and 10 MV FFF, respectively.Conclusions: We constructed an MC‐based QA system for evaluating patient treatment plans and evaluated its feasibility in clinical practice. We observed robust agreement between dose calculations from our QA system and measurements for VMAT plans. Our QA system could be useful in other clinical settings, such as small‐field SRS procedures or analyses of secondary cancer risk, for which dose calculations using TPS are difficult to verify.Keywords
Funding Information
- National Research Foundation
- Ministry of Education (NRF‐2016R1A6A3A11931787)
This publication has 45 references indexed in Scilit:
- Monte Carlo simulation on heterogeneous distributed systems: A computing framework with parallel merging and checkpointing strategiesFuture Generation Computer Systems, 2013
- American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) Practice Guideline for Intensity-modulated Radiation Therapy (IMRT)American Journal of Clinical Oncology, 2012
- Monte Carlo simulation of the dynamic micro-multileaf collimator of a LINAC Elekta Precise using PENELOPEPhysics in Medicine & Biology, 2011
- Monte Carlo evaluation of RapidArc™ oropharynx treatment planning strategies for sparing of midline structuresPhysics in Medicine & Biology, 2010
- Stereotactic body radiation therapy: The report of AAPM Task Group 101Medical Physics, 2010
- Measurement of depth-dose of linear accelerator and simulation by use of Geant4 computer codeReports of Practical Oncology and Radiotherapy, 2010
- Comparison between PENELOPE and electron Monte Carlo simulations of electron fields used in the treatment of conjunctival lymphomaPhysics in Medicine & Biology, 2009
- Monte Carlo simulation of RapidArc radiotherapy deliveryPhysics in Medicine & Biology, 2008
- Monte Carlo simulation of large electron fieldsPhysics in Medicine & Biology, 2008
- A 3D photon superposition/convolution algorithm and its foundation on results of Monte Carlo calculations*Physics in Medicine & Biology, 2005