Abstract
Background: The safety of performing magnetic resonance imaging (MRI) in patients with spinal cord stimulation (SCS) systems needs to be documented. A prospective in vivo study in patients with SCS, exploring the changes produced by MRI and the associated side effects, was performed. Methods: After ethics committee approval and patient consent, 31 consecutive patients with SCS at different spinal levels requiring a scheduled MRI evaluation were studied during an 18-month period. All MRIs were performed with a 1.5-T clinical use magnet and a specific absorption rate of no more than 0.9 W/kg. Frequency tables were used for the descriptive study, whereas comparative evaluations were made with the chi-square test for qualitative variables and single-factor analysis of variance for quantitative variables. Results: The mean patient age was 49 +/- 9.5 yr; 67.7% were women (n = 21), and 32.3% were men (n = 10). None of the patients experienced hemodynamic, respiratory, or neurologic alterations. Reported changes were as follows: increased temperature in the generator's area (n = 2, 6.5%); increased in the intensity of the stimulation (n = 1, 3.2%); impedance greater than 4,000 Omega on several of the electrodes in the leads (n = 1, 3.2%); telemetry not possible (n = 2, 6.5%). Radiologic evaluation after MRI revealed no spatial displacements of the SCS leads in any case. Conclusion: Under the conditions of the described protocol, MRI in patients with SCS systems resulted in few complications. None of the recorded problems were serious, and in no case were patients harmed or the systems reprogrammed. Maximum patient satisfaction was reported in all cases.Keywords
This publication has 18 references indexed in Scilit:
- Is Magnetic Resonance Imaging Safe for Patients with Neurostimulation Systems Used for Deep Brain Stimulation?Neurosurgery, 2005
- Bilateral neurostimulation systems used for deep brain stimulation: in vitro study of MRI-related heating at 1.5 T and implications for clinical imaging of the brainMagnetic Resonance Imaging, 2005
- In vivo heating of pacemaker leads during magnetic resonance imagingEuropean Heart Journal, 2004
- Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heatingJournal of Magnetic Resonance Imaging, 2004
- Neurostimulation System Used for Deep Brain Stimulation (DBS)Investigative Radiology, 2004
- Magnetic Resonance Imaging and Deep Brain StimulationNeurosurgery, 2002
- Magnetic resonance safety update 2002: Implants and devicesJournal of Magnetic Resonance Imaging, 2002
- Neurostimulation systems for deep brain stimulation: In vitro evaluation of magnetic resonance imaging–related heating at 1.5 teslaJournal of Magnetic Resonance Imaging, 2002
- Thalamic stimulation and functional magnetic resonance imaging: localization of cortical and subcortical activation with implanted electrodesJournal of Neurosurgery, 1999
- Magnetic Resonance Imaging with Implanted Neurostimulators: An In Vitro and In Vivo StudyNeurosurgery, 1999