Optimization of spin‐lock times in T1ρ mapping of knee cartilage: Cramér‐Rao bounds versus matched sampling‐fitting
- 4 November 2021
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 87 (3) , 1418-1434
Abstract
Purpose: To compare different optimization approaches for choosing the spin‐lock times (TSLs), in spin‐lattice relaxation time in the rotating frame (T1ρ) mapping.Methods: Optimization criteria for TSLs based on Cramér‐Rao lower bounds (CRLB) are compared with matched sampling‐fitting (MSF) approaches for T1ρ mapping on synthetic data, model phantoms, and knee cartilage. The MSF approaches are optimized using robust methods for noisy cost functions. The MSF approaches assume that optimal TSLs depend on the chosen fitting method. An iterative non‐linear least squares (NLS) and artificial neural networks (ANN) are tested as two possible T1ρ fitting methods for MSF approaches.Results: All optimized criteria were better than non‐optimized ones. However, we observe that a modified CRLB and an MSF based on the mean of the normalized absolute error (MNAE) were more robust optimization approaches, performing well in all tested cases. The optimized TSLs obtained the best performance with synthetic data (3.5–8.0% error), model phantoms (1.5–2.8% error), and healthy volunteers (7.7–21.1% error), showing stable and improved quality results, comparing to non‐optimized approaches (4.2–13.3% error on synthetic data, 2.1–6.2% error on model phantoms, 9.8–27.8% error on healthy volunteers).Conclusion: A modified CRLB and the MSF based on MNAE are robust optimization approaches for choosing TSLs in T1ρ mapping. All optimized criteria allowed good results even using rapid scans with two TSLs when a complex‐valued fitting is done with iterative NLS or ANN.Keywords
Funding Information
- National Institute of Biomedical Imaging and Bioengineering (P41‐EB017183)
- National Institute of Arthritis and Musculoskeletal and Skin Diseases (R01‐AR068966, R01‐AR076328‐01A1, R01‐AR076985‐01A1, R01‐AR078308‐01A1, R21‐AR075259‐01A1)
This publication has 46 references indexed in Scilit:
- Multipeak fat‐corrected complex R2* relaxometry: Theory, optimization, and clinical validationMagnetic Resonance in Medicine, 2013
- Knee Articular Cartilage Damage in Osteoarthritis: Analysis of MR Image Biomarker Reproducibility in ACRIN-PA 4001 Multicenter TrialRadiology, 2011
- Sparse MRI: The application of compressed sensing for rapid MR imagingMagnetic Resonance in Medicine, 2007
- Stretched exponential relaxation arising from a continuous sum of exponential decaysPhysical Review B, 2006
- An introduction to coil array design for parallel MRINMR in Biomedicine, 2006
- Optimization of diffusion measurements using Cramer‐Rao lower bound theory and its application to articular cartilageMagnetic Resonance in Medicine, 2003
- Proteoglycan‐induced changes in T1ρ‐relaxation of articular cartilage at 4TMagnetic Resonance in Medicine, 2001
- The stability of nonlinear least squares problems and the Cramer-Rao boundIEEE Transactions on Signal Processing, 2000
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- The Conjugate Gradient Method and Trust Regions in Large Scale OptimizationSIAM Journal on Numerical Analysis, 1983