Image Noise and Liver Lesion Detection With MDCT: A Phantom Study

Authors:

Abstract
OBJECTIVE. The purpose of this study was to determine the upper limit of noise for detection of small low-contrast lesions in a liver phantom. MATERIALS AND METHODS. A CT liver phantom containing 21 low-contrast, low-attenuation, circular simulated lesions ranging in size from 2.4 to 10 mm was scanned 23 times at different tube current ranges (varying noise index) on a 64-MDCT scanner with automatic tube current modulation. The attenuation of the simulated lesions was 20 HU less than that of the liver-equivalent background. Three radiologists independently reviewed the resultant CT images, which contained either a low-contrast lesion or no lesion and scored certainty of lesion detection using a 4-point Likert scale. Overall performance was evaluated by sensitivity analysis with receiver operator curve and area under the curve (Az) computation for ranges of noise index. RESULTS. The reviewers achieved 100% sensitivity with a noise index of 15 or less for lesions measuring 6.3–10.0 mm (Az = 0.96). Increasin...