Filter-based ultralow-frequency Raman measurement down to 2 cm−1 for fast Brillouin spectroscopy measurement

Authors:

Abstract
Simultaneous Stokes and anti-Stokes ultralow-frequency (ULF) Raman measurement down to ∼2 cm−1 or 60 GHz is realized by a single-stage spectrometer in combination with volume-Bragg-grating-based notch filters. This system reveals its excellent performance by probing Brillouin signal of acoustic phonons in silicon, germanium, gallium arsenide, and gallium nitride. The deduced sound velocity and elastic constants are in good accordance with previous results determined by various methods. This system can shorten the integration time of the Brillouin signal with a good signal-to-noise ratio by more than 2000-fold compared to a Fabry-Perot interferometer (FPI). This study shows how a filter-based ULF Raman system can be used to reliably achieve Brillouin spectroscopy for condensed materials with high sensitivity and high signal-to-noise ratio, stimulating fast Brillouin spectrum measurements to probe acoustic phonons in semiconductors.
Funding Information
  • National Key Research and Development Program of China (2016YFA0301204)
  • National Natural Science Foundation of China (11225421, 11434010, 11474277)

This publication has 49 references indexed in Scilit: