Abstract
The Co-based special wire alloys 35N LT and MP35N are used for biomedical applications. The corrosion resistance of the two alloys, chemical passivation, and metal release of 35N LT were studied by electrochemical measurements, surface analysis, and solution analysis. The results show no significant difference in the corrosion resistance between 35N LT and MP35N. Preferential dissolution of Co and Ni occurs during chemical passivation of 35N LT in an HNO3HNO3 solution, which results in enrichment of oxidized Cr and depletion of oxidized Co in the passive film. The passivation treatment leads to an increased corrosion resistance in a phosphate-buffered saline solution, even with addition of 100mM100mM H2O2H2O2 simulating an inflammatory response. A further effect of the passivation treatment is a significantly reduced metal release, especially Co, from 35N LT during exposure to the synthetic biological medium.Keywords
This publication has 12 references indexed in Scilit:
- In vivo biostability of polysiloxane polyether polyurethanes: Resistance to metal ion oxidationJournal of Biomedical Materials Research Part A, 2006
- Electrochemical Behavior of Stimulation∕Sensing Materials for Pacemaker Electrode ApplicationsJournal of the Electrochemical Society, 2005
- Biodegradation of polyether polyurethane inner insulation in bipolar pacemaker leadsJournal of Biomedical Materials Research, 2001
- Eis and XPS study of surface modification of 316LVM stainless steel after passivationCorrosion Science, 1998
- Electrochemical impedance spectroscopy study of the passive oxide film on titanium for implant applicationElectrochimica Acta, 1996
- Hydrogen peroxide toward enhanced oxide growth on titanium in PBS solution: Blue coloration and clinical relevanceJournal of Biomedical Materials Research, 1996
- Polyurethane Elastomer BiostabilityJournal of Biomaterials Applications, 1995
- Comparison of electrical transients and corrosion responses of pulsed MP35N and 316LVM electrodesAnnals of Biomedical Engineering, 1994
- Comparison of 316LVM and MP35N alloys as charge injection electrodesJournal of Biomedical Materials Research, 1994
- Implantable pacing lead technologyIEEE Engineering in Medicine and Biology Magazine, 1990