Comparison of PMT‐based TF64 and SiPM‐based Vereos PET/CT systems for 90Y imaging and dosimetry optimization: A quantitative study
Authors:
- 8 August 2022
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 49 (12) , 7567-7582
Abstract
Background: Selective internal radiotherapy based on transarterial radioembolization (TARE) with yttrium‐90 (90Y) microspheres is an established treatment for primary or metastatic liver disease.Purpose: The objective of this work is to optimize the dosimetry of patients treated with 90Y TARE, using positron emission tomography (PET) images.Methods: The NEMA 2012 PET phantom was filled with nearly 3.9 GBq of 90Y activity and acquired at days 0, 3, 5, 7, and 9 on a classic time‐of‐flight PET/computed tomography (CT) scanner (Philips TF64) and on a silicon photomultiplier (SiPM)‐based PET/CT scanner (Philips Vereos). Acquisitions were carried on following the guidelines proposed in a previously published multicentric trial and images were reconstructed by varying and combining the available parameters. Comparisons were performed to identify the best set(s) of parameters leading to the most accurate 90Y‐PET image(s), in terms of activity distribution. Then, for both scanners, the best images were analyzed with Simplicit90Y, a personalized dosimetry software using multicompartmental Medical Internal Radiation Dose model. The comparison between measured and true doses allowed to identify the image granting the most consistent dose estimations and, therefore, to designate the set of parameters to be applied on patients’ data for the reconstruction of optimized clinical images. Posttreatment dosimetry of four patients was then realized with Simplicit90Y using optimized imaging datasets.Results: Based on activity distribution comparisons and dose estimations over phantom and patients data, the SiPM‐based PET/CT system appeared more suitable than the photomultiplier tube‐based TF64 for 90Y‐PET imaging. With the SiPM‐based PET/CT system, reconstructed images with a 2‐mm voxel size combined with the application of the point spread function correction led to the most accurate results for quantitative 90Y measures.Conclusions: For the SiPM‐based PET/CT scanner, an optimized set of reconstruction parameters has been identified and applied on patients’ data in order to generate the most accurate image to be used for an improved personalized 90Y‐PET dosimetry, ensuring a reliable evaluation of the delivered doses.Keywords
This publication has 22 references indexed in Scilit:
- Radioembolization and the Dynamic Role of 90Y PET/CTFrontiers in Oncology, 2014
- Effect of Varying Number of OSEM Subsets on PET Lesion DetectabilityJournal of Nuclear Medicine Technology, 2013
- Emission of β+ Particles Via Internal Pair Production in the 0+ – 0+ Transition of 90Zr: Historical Background and Current Applications in Nuclear Medicine ImagingAtoms, 2013
- Dosimetry Based on 99mTc-Macroaggregated Albumin SPECT/CT Accurately Predicts Tumor Response and Survival in Hepatocellular Carcinoma Patients Treated with 90Y-Loaded Glass Microspheres: Preliminary ResultsJournal of Nuclear Medicine, 2012
- Quantitative Evaluation of Scintillation Camera Imaging Characteristics of Isotopes Used in Liver RadioembolizationPLOS ONE, 2011
- Partial-Volume Effect in PET Tumor ImagingJournal of Nuclear Medicine, 2007
- Treatment of Unresectable Hepatocellular Carcinoma with Use of 90Y Microspheres (TheraSphere): Safety, Tumor Response, and SurvivalJournal of Vascular and Interventional Radiology, 2005
- Yttrium-90 Microsphere Therapy for Hepatic Malignancy: Devices, Indications, Technical Considerations, and Potential ComplicationsRadioGraphics, 2005
- Clinical evaluation of the partition model for estimating radiation doses from yttrium-90 microspheres in the treatment of hepatic cancerEuropean Journal of Nuclear Medicine and Molecular Imaging, 1997
- Accelerated image reconstruction using ordered subsets of projection dataIEEE Transactions on Medical Imaging, 1994