Detection of Approximal Caries with a New Laser Fluorescence Device
Authors:
- A. LussiALA. Lussi
- A. HackAHA. Hack
- I. HugIHI. Hug
- H. HeckenbergerHHH. Heckenberger
- B. MegertBMB. Megert
- H. StichHSH. Stich
- 1 February 2006
- journal article
- research article
- Published by S. Karger AG in Caries Research
- Vol. 40 (2) , 97-103
Abstract
The laser device DIAGNOdent developed for the detection of occlusal caries has limited value on approximal surfaces. The aim of this study was to develop and to test a new laser fluorescence (LF) device for the detection of approximal caries. Light with a wavelength of 655 nm was transported to the approximal surface using two different sapphire fibre tips. Seventy-five teeth were selected from a pool of extracted permanent human molars, frozen at –20°C until use. Before being measured, they were defrosted, cleaned and calculus was removed with a scaler. The molars were set in blocks simulating the contact area of adults. Bitewing radiographs were obtained using Kodak Insight films. After two independent assessments with the new LF device, the teeth were histologically prepared, and assessed for caries extension. Using the laser, specificity values for D1 threshold (outer half of enamel), D2 threshold (inner half of enamel), D3 threshold (dentine) ranged between 0.81 and 0.93, sensitivity between 0.84 and 0.92 with no difference between the two tips. Bitewing radiography showed an inferior performance compared to LF (p < 0.05). Intraex aminer reproducibility was high (kappa >0.74). The new LF system might be a useful additional tool in detecting approximal caries. Because of its good reproducibility, it could be used to monitor caries regression or progression on approximal surfaces.Keywords
This publication has 7 references indexed in Scilit:
- A systematic review of the performance of a laser fluorescence device for detecting cariesThe Journal of the American Dental Association, 2004
- Improving Clinical Visual Detection—Potential for Caries Clinical TrialsJournal of Dental Research, 2004
- Bitewing and Digital Bitewing Radiography for Detection of Caries LesionsJournal of Dental Research, 2004
- In vitro study of caries detection through sound dentin using a laser fluorescence device, DIAGNOdentEuropean Journal of Oral Sciences, 2003
- Clinical performance of a laser fluorescence device for detection of occlusal caries lesionsEuropean Journal of Oral Sciences, 2001
- Performance and Reproducibility of a Laser Fluorescence System for Detection of Occlusal Caries in vitroCaries Research, 1999
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986