Determination of the solid fat content of commercial fats by pulsed nuclear magnetic resonance
Authors:
- 1 March 1978
- journal article
- research article
- Published by Wiley in Journal of the American Oil Chemists' Society
- Vol. 55 (3) , 328-331
Abstract
A method and instrumentation for measuring the solid fat content is reported that is both accurate and precise. It involves using transient nuclear magnetic resonance (NMR) measurements for determining the percentage of solids in commercial shortenings and hydrogenated oils at selected temperatures. The incorporation of a tempering step at 26.7 C for all samples before measurement has improved the precision of the solids content actually measured (±0.2% solids) which approximates that of dilatometry. Duplicate measurements are not required to obtain this precision. The instrument is equipped with six 10‐mm sample holders in combination with a precise variable temperature accessory system which eliminates the temperature difference between the sample and sample holder. This improved and exact temperature conditioning of samples provides better sample stability and easier handling for routine conditions of analysis. A single temperature result can be made in less than 1 hr and typical 5 temperature results obtained in 2.5 hr. Our work also indicates that tempering does influence results, the net effect being to decrease the amount of solids at temperatures less than the tempering temperature. In comparing the pulsed NMR measured solids with these measured by dilatometry, differences between methods of measurement are minimized when samples have had the same tempering and temperature history. This method provides flexibility, speed, and increased sample throughput of up to 60 samples/day. The self‐contained equipment requires only 9 sq ft of space and is ready for measurements within 45 min after start‐up.Keywords
This publication has 6 references indexed in Scilit:
- EVALUATION OF SEVERAL PULSED NMR TECHNIQUES FOR SOLIDS‐IN‐FAT DETERMINATION OF COMMERCIAL FATSJournal of Food Science, 1977
- Fully automated determination of solid fat content by pulsed NMRJournal of the American Oil Chemists' Society, 1974
- Report of the instrumental techniques committee AOCS, 1971–72Journal of the American Oil Chemists' Society, 1972
- Measurement of the Solids Content of Fats by Nuclear Magnetic ResonanceCanadian Institute of Food Science and Technology Journal, 1972
- A nuclear magnetic resonance study of the liquid/solid content of margarine fatJournal of the American Oil Chemists' Society, 1960