A comparison of the forces exerted during laryngoscopy using disposable and non‐disposable laryngoscope blades
Authors:
Open Access
- 1 August 2003
- journal article
- clinical trial
- Published by Wiley in Anaesthesia
- Vol. 58 (9) , 869-873
Abstract
Summary The concern that re‐usable equipment may be a source of cross‐infection has prompted the increased use of disposable laryngoscope blades. We conducted a study investigating the duration of laryngoscopy and the peak force generated using various laryngoscope blades. Five blades were studied: one reusable Macintosh 3 blade, one disposable metal blade and three plastic disposable blades. Sixty anaesthetists performed laryngoscopy on a mannequin using each of the five blades presented in a random order. This was then repeated with a rigid collar applied to the mannequin. The mean force applied with the collar off with the metal blades was 32.8 and 30.8 N, and with the plastic blades 37.3, 39.6 and 41.5 N, respectively (p < 0.0001). The mean force applied with the collar on with the metal blades was 30.5 and 32.5 N, and with the plastic blades 35.5, 34.9 and 31.4 N, respectively (p < 0.0001). The mean duration of laryngoscopy with the collar off for the metal blades was 5.6 and 5.4 s, and with the plastic blades 10, 7.2 and 7.5 s (p < 0.0001). The duration of laryngoscopy with the collar applied for the metal blades was 7.1 and 7.5 s, and with the plastic blades 11.8, 9.7 and 9.0 s (p < 0.0001). The use of plastic blades results in both greater peak force and duration of laryngoscopy.Keywords
This publication has 13 references indexed in Scilit:
- Presence of protein deposits on 'cleaned' re-usable anaesthetic equipment. apparatusAnaesthesia, 2001
- Presence of protein deposits on ‘cleaned’ re‐usable anaesthetic equipmentAnaesthesia, 2001
- A comparison of the forces exerted during laryngoscopyAnaesthesia, 1996
- A new variant of Creutzfeldt-Jakob disease in the UKThe Lancet, 1996
- Force and Torque Vary Between Laryngoscopists and Laryngoscope BladesAnesthesia & Analgesia, 1996
- Does experience influence the forces exerted on maxillary incisors during laryngoscopy? A manikin study using the Macintosh laryngoscopeCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1995
- Forces acting on the maxillary incisor teeth during laryngoscopy using the Macintosh laryngoscopeAnaesthesia, 1994
- Measurement of forces during laryngoscopyAnaesthesia, 1992
- Force Applied During Tracheal IntubationAnesthesia & Analgesia, 1992
- Difficult tracheal intubation in obstetricsAnaesthesia, 1984