Calibration of a pyroelectric detector at 106 µm with the National Institute of Standards and Technology high-accuracy cryogenic radiometer
Authors:
- T. R. GentileTGT. R. Gentile
- J. M. HoustonJHJ. M. Houston
- G. EppeldauerGEG. Eppeldauer
- A. L. MigdallAMA. L. Migdall
- C. L. CromerCCC. L. Cromer
- 1 June 1997
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 36 (16) , 3614-3621
Abstract
The National Institute of Standards and Technology (NIST) is establishing an infrared detector calibration facility to improve radiometric standards at infrared wavelengths. The absolute response of the cryogenic bolometer that serves as the transfer standard for this facility is being linked to the NIST high- accuracy cryogenic radiometer (HACR) at a few laser wavelengths. At the 10.6-microm CO(2) laser line, this link is being established through a pyroelectric detector that has been calibrated against the HACR. We describe the apparatus, methods, and uncertainties for the calibration of this pyroelectric detector.Keywords
This publication has 4 references indexed in Scilit:
- National Institute of Standards and Technology high-accuracy cryogenic radiometerApplied Optics, 1996
- Deposition and characterization of far-infrared absorbing gold black filmsApplied Optics, 1993
- Characterization of a High Sensitivity Composite Silicon BolometerMetrologia, 1993
- Low-temperature behavior of spontaneous polarization in LiNband LiTaPhysical Review B, 1976