Electrically Calibrated Pyroelectric Optical-Radiation Detector
Authors:
- 1 October 1973
- journal article
- Published by Optica Publishing Group in Applied Optics
- Vol. 12 (10) , 2494-2500
Abstract
An electrically calibrated optical detector has been developed using a pyroelectric response of the plastic, polyvinylfluoride. An in-depth look at the modulation frequency response was performed to substantiate the equivalence of the optical and electrical inputs, indicate the optimum structure and allow for a clearer understanding of the device limitations. The experimental results of the dynamic range, linearity, uniformity, and detectivity confirm the device’s utility.Keywords
This publication has 7 references indexed in Scilit:
- Sinusoidally modulated heat flow and the pyroelectric effectInfrared Physics, 1972
- A reference calorimeter for laser energy measurementsJournal of Research of the National Bureau of Standards Section A: Physics and Chemistry, 1972
- High D* Pyroelectric Polyvinylfluoride DetectorsApplied Physics Letters, 1971
- Thermal analysis of pyroelectric detectorsInfrared Physics, 1970
- Dynamic Method for Measuring the Pyroelectric Effect with Special Reference to Barium TitanateJournal of Applied Physics, 1956
- A new absolute radiometerPhysica, 1951
- Studies of instruments for measuring radiant energy in absolute value: an absolute thermopileBulletin of the Bureau of Standards, 1916