A novel method of temperature compensation for piezoresistive microcantilever-based sensors
Authors:
- 1 March 2012
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 83 (3) , 035002
Abstract
Microcantilever with integrated piezoresistor has been applied to in situ surface stress measurement in the field of biochemical sensors. It is well known that piezoresistive cantilever-based sensors are sensitive to ambient temperature changing due to highly temperature-dependent piezoresistive effect and mismatch in thermal expansion of composite materials. This paper proposes a novel method of temperature drift compensation for microcantilever-based sensors with a piezoresistive full Wheatstone bridge integrated at the clamped ends by subtracting the amplified output voltage of the reference cantilever from the output voltage of the sensing cantilever through a simple temperature compensating circuit. Experiments show that the temperature drift of microcantilever sensors can be significantly reduced by the method.Keywords
Funding Information
- National Natural Science Foundation of China (51075377, 61076110)
This publication has 23 references indexed in Scilit:
- Design and Fabrication of Piezoresistive Based Encapsulated Poly-Si Cantilevers for Bio/chemical SensingPhysics Procedia, 2011
- AFM probes fabricated with masked–maskless combined anisotropic etching and p+surface dopingJournal of Micromechanics and Microengineering, 2005
- Cantilever transducers as a platform for chemical and biological sensorsReview of Scientific Instruments, 2004
- Microcantilevers integrated with heaters and piezoelectric detectors for nano data-storage applicationApplied Physics Letters, 2003
- Micromachined piezoresistive cantilever array with integrated resistive microheater for calorimetry and mass detectionJournal of Vacuum Science & Technology A, 2001
- Micromechanical cantilever-based biosensorsSensors and Actuators B: Chemical, 2001
- A microcantilever-based alcohol vapor sensor-application and response modelApplied Physics Letters, 2000
- Fabrication of improved piezoresistive silicon cantilever probes for the atomic force microscopeSensors and Actuators A: Physical, 1997
- Micromachined silicon sensors for scanning force microscopyJournal of Vacuum Science & Technology B: Microelectronics and Nanometer Structures Processing, Measurement, and Phenomena, 1991
- Microfabrication of cantilever styli for the atomic force microscopeJournal of Vacuum Science & Technology A, 1990