Physical performance of a long axial field‐of‐view PET scanner prototype with sparse rings configuration: A Monte Carlo simulation study
- 27 January 2020
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 47 (4) , 1949-1957
Abstract
Purpose: There is a growing interest in extending the axial fields‐of‐view (AFOV) of PET scanners. One major limitation for the widespread clinical adoption of such systems is the multifold increase in the associated material costs. In this study, we propose a cost‐effective solution to extend the PET AFOV using a sparse detector rings configuration. The corresponding physical performance was validated using Monte Carlo simulations.Methods: Monte Carlo model of the Siemens BiographTM mCT PET/CT, with a 21.8 cm AFOV and a set of compact rings of LSO crystals was developed as a gold standard. The mCT configuration was then modified by interleaving the LSO crystals in the axial direction within each detector block with 4 mm physical gaps (equivalent to the LSO crystal axial dimension) thus extending the AFOV to 43.6 cm (Ex‐mCT). The physical performances of the two MC models were assessed and then compared using NEMA NU 2‐2007 standards.Results: Ex‐mCT showed <0.2 mm difference in transaxial spatial resolution, and, 0.8 mm and 0.3 mm deterioration in axial spatial resolution, compared to the mCT, at 1 and 10 cm off‐center of the transaxial field‐of‐view respectively. The system sensitivities for the mCT and Ex‐mCT models were 9.4 ± 0.2 and 10.75 ± 0.2 cps/kBq respectively. The higher sensitivity of Ex‐mCT was due to four additional detector rings required to double the mCT AFOV. PET images of the NEMA Image Quality (IQ) phantom showed no artifacts due to detector rings sparsity, and all spheres were visible in both configurations. Ex‐mCT achieved percent contrast recoveries within 5.6% of those of the mCT for all spheres and a maximum of 36% higher background variability at the center of the AFOV. The Ex‐mCT, however, showed a more uniform noise distribution over an axial range of almost twice the length of the mCT AFOV.Conclusions: Using the proposed sparse detector‐ring configuration, the AFOV of current generation PET systems can be doubled while maintaining the original number and volume of detector crystal elements, and without jeopardizing the system’s overall physical performance. Despite an increase in the noise level, the Ex‐mCT exhibited an improved noise uniformity.Keywords
This publication has 36 references indexed in Scilit:
- Optimal whole-body PET scanner configurations for different volumes of LSO scintillator: a simulation studyPhysics in Medicine & Biology, 2012
- STIR: software for tomographic image reconstruction release 2Physics in Medicine & Biology, 2012
- Physical and clinical performance of the mCT time-of-flight PET/CT scannerPhysics in Medicine & Biology, 2011
- Advances in Imaging the Brain–Gut Axis: Functional Gastrointestinal DisordersGastroenterology, 2011
- An Assessment of the Impact of Incorporating Time-of-Flight Information into Clinical PET/CT ImagingJournal of Nuclear Medicine, 2010
- Simulation studies of a new ‘OpenPET’ geometry based on a quad unit of detector ringsPhysics in Medicine & Biology, 2009
- A proposal of an open PET geometryPhysics in Medicine & Biology, 2008
- Deep-Inspiration Breath-Hold PET/CT: Clinical Findings with a New Technique for Detection and Characterization of Thoracic LesionsJournal of Nuclear Medicine, 2007
- Performance evaluation of the new whole-body PET/CT scanner: Discovery STEuropean Journal of Nuclear Medicine and Molecular Imaging, 2004
- Analytic 3D image reconstruction using all detected eventsIEEE Transactions on Nuclear Science, 1989