A Low Stray Light Raman Microprobe Using Optical Fibers and GRIN Lenses
Authors:
- 1 September 1991
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 45 (8) , 1302-1304
Abstract
The demonstrated Raman microprobe uses optical fibers and GRIN lenses to focus an exciting laser and record Raman scattering. Due to the small size of the optical components, the two parts are spatially separated; therefore much specular light or diffused light is removed from the collected light. Five collecting channels are used to increase the collecting efficiency. Experiments show that the stray light is efficiently reduced and the intensity of the Raman signal is more than ten times higher than the intensity of the spectrum obtained with a conventional Raman microscope. An exciting spot of less than 9 μm in diameter is obtained.Keywords
This publication has 5 references indexed in Scilit:
- Micro-Raman Spectrum of Triuranium Octoxide, U 3 O 8Applied Spectroscopy, 1988
- Microprobe accessory for laser raman spectrometersOptics & Laser Technology, 1987
- Raman Microprobe and Fourier Transform-Infrared Microsampling Studies of the Microstructure of GallstonesApplied Spectroscopy, 1987
- Raman microprobe analysis of preforms and optical fibersApplied Optics, 1984
- Raman microprobe and microscope with laser excitationJournal of Raman Spectroscopy, 1975