Efficacy and safety of capsular bending ring implantation to prevent posterior capsule opacification
Authors:
- Rupert MenapaceRMRupert Menapace
- Stefan SacuSSStefan Sacu
- Michael GeorgopoulosMGMichael Georgopoulos
- Oliver FindlOFOliver Findl
- Georg RainerGRGeorg Rainer
- Okihiro NishiONOkihiro Nishi
- 1 August 2008
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 34 (8) , 1318-1328
Abstract
To determine whether a capsular bending ring (CBR) with a rectangular cross-section and sharp edges moves the barrier to the very equator and avoids contact between the capsulorhexis and optic to prevent posterior capsule opacification (PCO) and anterior capsule fibrosis. Department of Ophthalmology, Medical University of Vienna, Vienna, Austria. A 0.7 mm high, open poly(methyl methacrylate) CBR was implanted in 60 eyes (patients) in a prospective randomized intraindividual trial. The impact of additional CBR implantation on PCO and anterior capsule fibrosis was compared to that of intraocular lens (IOL) implantation alone using objective scoring. No CBR-related surgical complications occurred. The objective PCO score and area were statistically significantly reduced in the CBR group. In patients with complete follow-up, the mean PCO score (scale 1 to 10) at 1, 2, and 3 years was 0.8, 1.7, and 2.1, respectively, in the CBR group and 2.6, 3.9, and 4.6, respectively, in the no-CBR group. The number of quadrants affected by PCO was 0.9, 1.5, and 1.8 versus 3.2, 3.8, and 3.8. Barrier failures with the CBR were caused by the inherent slight edge blunting and occasional eyelet gaping. Laser capsulotomies were performed in the no-CBR group only. Capsule stress folds and fibrotic anterior capsule opacification were also greatly reduced. The best corrected visual acuity was better in the CBR group. Capsular bending ring implantation was an effective and safe adjunct to in-the-bag IOL fixation. With improvements in technology and design securing exquisitely sharp edges and circumferential capsular bending independent of the capsular bag diameter, this concept has the potential to prevent PCO and anterior capsule fibrosis.Keywords
This publication has 19 references indexed in Scilit:
- Long-term Effect of Optic Edge Design in a Silicone Intraocular Lens on Posterior Capsule OpacificationAmerican Journal of Ophthalmology, 2007
- Contact inhibition of migrating lens epithelial cells at the capsular bend created by a sharp-edged intraocular lens after cataract surgeryJournal of Cataract & Refractive Surgery, 2007
- Long-term Effect of Sharp Optic Edges of a Polymethyl Methacrylate Intraocular Lens on Posterior Capsule OpacificationOphthalmology, 2005
- Evaluating and defining the sharpness of intraocular lensesJournal of Cataract & Refractive Surgery, 2005
- Long-term effect of optic edge design in an acrylic intraocular lens on posterior capsule opacificationJournal of Cataract & Refractive Surgery, 2005
- Cataract surgical problemJournal of Cataract & Refractive Surgery, 2005
- Long-term efficacy of adding a sharp posterior optic edge to a three-piece silicone intraocular lens on capsule opacification: Five-year results of a randomized studyAmerican Journal of Ophthalmology, 2005
- Closed foldable capsular ringsJournal of Cataract & Refractive Surgery, 2005
- Inhibition of lens epithelial cell migration at the intraocular lens optic edgeJournal of Cataract & Refractive Surgery, 2003
- Efficacy of Equator Rings in an Experimental Rabbit StudyArchives of Ophthalmology (1950), 1995