Abstract
Two charge multiplexed readout schemes for Position Sensitive Photomultiplier Tubes (PSPMTs) have been developed and evaluated. A two-stage split-charge method is compared to standard positional charge division. The resulting electronic circuits are highly compact and will be initially incorporated into a small hand-held nuclear gamma ray imager for cancer staging. Electrical tests on the readout electronics are performed and intrinsic scintillation camera performance is evaluated using a 23/spl times/23 array of pixilated 2/spl times/2/spl times/3 mm/sup 3/ LSO crystals coupled to a H8500 flat-panel PSPMT.Keywords
This publication has 4 references indexed in Scilit:
- Evaluation of performance of dedicated, compact scintillation cameras2002
- Multi-PSPMT scintillation cameraIEEE Transactions on Nuclear Science, 1999
- Performance characteristics of a compact position-sensitive LSO detector moduleIEEE Transactions on Medical Imaging, 1998
- Simple charge division readouts for imaging scintillator arrays using a multi-channel PMTIEEE Transactions on Nuclear Science, 1996