Multi-parameter analysis of the effects on hydraulic performance and hemolysis of blood pump splitter blades
Open Access
- 1 May 2014
- journal article
- research article
- Published by SAGE Publications in Advances in Mechanical Engineering
- Vol. 12 (5)
Abstract
The splitter blade can effectively optimize pump performance, but there is still insufficient research in blood pumps that cover both hydraulic and hemolysis performance. Thus, the aim of this study was to investigate the effect of key factors related to splitter blade on the performance and flow field of axial flow blood pump. In this study, the number of splitter blades, the axial length, and the circumferential offset were chosen as three objects of study. An analysis of the flow field and performance of the pump by orthogonal array design using computational fluid mechanics was carried out. A set of hydraulic and particle image velocimetry experiments of the model pumps were performed. The result showed that the pump had greater hydraulic performance without sacrificing its hemolytic performance when it had two splitter blades, the axial length ratio was 0.6, and the circumferential offset was 15°. Based on these reference data, the splitter blade may contribute to greater hydraulic performance of the pump and cause no side effect on the velocity distribution of the flow field. This finding provides an effective method for the research, development, and application of structural improvement of the axial flow blood pump.Keywords
Funding Information
- Youth Science Fund Project (51475477)
- Central South University (2018zzts021)
This publication has 25 references indexed in Scilit:
- In Vitro Assessment of the Apico Aortic Blood Pump: Anatomical Positioning, Hydrodynamic Performance, Hemolysis Studies, and Analysis in a Hybrid Cardiovascular SimulatorArtificial Organs, 2013
- Modeling of a Rotary Blood PumpArtificial Organs, 2013
- The simulation of multiphase flow field in implantable blood pump and analysis of hemolytic capabilityJournal of Hydrodynamics, 2013
- Unsteady internal flow conditions of mini-centrifugal pump with splitter bladesJournal of Thermal Science, 2013
- NUMERICAL ANALYSIS OF AN AXIAL BLOOD PUMP WITH DIFFERENT IMPELLER BLADE HEIGHTSJournal of Mechanics in Medicine and Biology, 2012
- 3D Particle image velocimetry test of inner flow in a double blade pump impellerChinese Journal of Mechanical Engineering, 2012
- Numerical Investigation on Hydrodynamics and Biocompatibility of a Magnetically Suspended Axial Blood PumpASAIO Journal, 2006
- The Hemolytic Characteristics of Monopivot Magnetic Suspension Blood Pumps With Washout HolesArtificial Organs, 2005
- Three‐Dimensional Numerical Prediction of Stress Loading of Blood Particles in a Centrifugal PumpArtificial Organs, 1995
- Model for a General Mechanical Blood Damage PredictionArtificial Organs, 1995