Copper filtration in pediatric digital X-ray imaging: its impact on image quality and dose
Authors:
- 23 March 2011
- journal article
- Published by Springer Nature in Radiological Physics and Technology
- Vol. 4 (2) , 148-155
Abstract
The effect of copper (Cu) filtration on image quality and dose in different digital X-ray systems was investigated. Two computed radiography systems and one digital radiography detector were used. Three different polymethylmethacrylate blocks simulated the pediatric body. The effect of Cu filters of 0.1, 0.2, and 0.3 mm thickness on the entrance surface dose (ESD) and the corresponding effective doses (EDs) were measured at tube voltages of 60, 66, and 73 kV. Image quality was evaluated in a contrast-detail phantom with an automated analyzer software. Cu filters of 0.1, 0.2, and 0.3 mm thickness decreased the ESD by 25–32%, 32–39%, and 40–44%, respectively, the ranges depending on the respective tube voltages. There was no consistent decline in image quality due to increasing Cu filtration. The estimated ED of anterior-posterior (AP) chest projections was reduced by up to 23%. No relevant reduction in the ED was noted in AP radiographs of the abdomen and pelvis or in posterior–anterior radiographs of the chest. Cu filtration reduces the ESD, but generally does not reduce the effective dose. Cu filters can help protect radiosensitive superficial organs, such as the mammary glands in AP chest projections.Keywords
This publication has 15 references indexed in Scilit:
- Digital radiography: The balance between image quality and required radiation doseEuropean Journal of Radiology, 2009
- Advances in Digital Radiography: Physical Principles and System OverviewRadioGraphics, 2007
- Evaluation of a software package for automated quality assessment of contrast detail images—comparison with subjective visual assessmentPhysics in Medicine & Biology, 2005
- Dose Reduction in Patients Undergoing Chest Imaging: Digital Amorphous Silicon Flat-Panel Detector Radiography Versus Conventional Film-Screen Radiography and Phosphor-Based Computed RadiographyAmerican Journal of Roentgenology, 2003
- Radiation Dose Reduction in Chest Radiography using a Flat-Panel Amorphous Silicon DetectorClinical Radiology, 2002
- Chest imaging with flat-panel detector at low and standard doses: comparison with storage phosphor technology in normal patientsEuropean Radiology, 2001
- Guidelines on Best Practice in the X-Ray Imaging of Children, By J.V. Cook, K. Shah, S. Pablot, K. Kyriou, A. Pettet, M. Fitzgerald, Queen Mary’s Hospital for Children.European Journal of Radiology, 2000
- Radiation dose reduction in paediatric fluoroscopy using added filtrationBritish Journal of Radiology, 1995
- The influence of tube filtration and potential on patient dose during X-ray examinationsPhysics in Medicine & Biology, 1988
- Filter materials for dose reduction in screen-film radiographyPhysics in Medicine & Biology, 1986