Field Evaluation of a Polymerase Chain Reaction-Based Nonisotopic Liquid Hybridization Assay for Malaria Diagnosis

Authors:
  • YSY. P. Shi
    Y. P. Shi
  • AOAggrey J. Oloo
    Aggrey J. Oloo
  • DBDavid A. Boriga
    David A. Boriga
  • BNBernard L. Nahlen
    Bernard L. Nahlen
  • WHWilliam A. Hawley
    William A. Hawley
  • BHBrian P. Holloway
    Brian P. Holloway
  • ALAltaf A. Lal
    Altaf A. Lal

Abstract
In a blind field evaluation of a nonisotopic liquid hybridization assay for detection of malaria parasites, 100 blood samples were tested from an area in which malaria is endemic; light microscopy was used as the reference test. Sensitivity, specificity, and positive and negative predictive values of the hybridization assay were 100%. One sample that was microscopy-negative and hybridization-positive was positive when reexamined. Another sample that was microscopy-positive and hybridization- negative was negative at reexamination. The detection limit of the test was ⩾0.0005% parasitemia. Four samples with mixed infections were misdiagnosed by microscopy as single-species infections. Four samples diagnosed as mixed infections by microscopy and single infection by the hybridization test had no evidence of a second Plasmodium species upon reexamination. The polymerase chain reaction - based nonisotopic liquid hybridization assay was better than conventional light microscopy in detecting low-grade parasite infection and offers an exceptional advantage for detecting mixed infections.