Sm:YAG optical pressure sensor to 180 GPa:Calibration and structural disorder
Authors:
- 20 June 1994
- journal article
- research article
- Published by AIP Publishing in Applied Physics Letters
- Vol. 64 (25) , 3386-3388
Abstract
We report ultrahigh pressure calibration and fluorescence intensity measurements of Sm:YAG to 180 GPa in a diamond anvil cell at room temperature. Several fluorescence emission bands excited by argon ion laser were followed with increasing pressure and it was observed that Y4 band was the most intense in the pressure regime of 100–180 GPa. The Y4 band is calibrated against the R band of ruby to 180 GPa. Irreversible changes in the line intensities above 100 GPa were observed in Sm:YAG and quenched to ambient pressure. Linewidths of the fluorescence bands of the quenched phase indicate structural disorder. In contrast to ruby pressure sensor, the Y4 emission band of Sm:YAG is easily detectable at ultrahigh pressures.Keywords
This publication has 12 references indexed in Scilit:
- Comparison of the pressure-induced frequency shift of Sm:YAG to the ruby and nitrogen vibron pressure scales from 6 to 820 K and 0 to 25 GPa and suggestions for use as a high-temperature pressure calibrantJournal of Applied Physics, 1992
- Optical transitions in diamond at ultrahigh pressuresNature, 1991
- Calibration of Sm:YAG as an alternate high-pressure scaleJournal of Applied Physics, 1990
- Pressure and temperature dependence of laser-induced fluorescence of Sm:YAG to 100 kbar and 700 °C and an empirical modelJournal of Applied Physics, 1990
- Pressure dependence of theandabsorption bands of ruby to 35 GPaPhysical Review B, 1990
- Temperature and pressure dependence of laser induced fluorescence in Sm:YAG—a new pressure calibrantHigh Pressure Research, 1989
- Model line-shape analysis for the ruby R lines used for pressure measurementJournal of Applied Physics, 1985
- Diamond Anvil Pressure Cell and Pressure Sensor for High-Temperature UseJapanese Journal of Applied Physics, 1982
- Specific volume measurements of Cu, Mo, Pd, and Ag and calibration of the ruby R1 fluorescence pressure gauge from 0.06 to 1 MbarJournal of Applied Physics, 1978
- An Optical Fluorescence System for Quantitative Pressure Measurement in the Diamond-Anvil CellReview of Scientific Instruments, 1973