Overset meshing in combination with novel blended weak-strong fluid-structure interactions for simulations of a translating valve in series with a second valve
Authors:
Open Access
- 18 April 2023
- journal article
- research article
- Published by Taylor & Francis in Computer Methods in Biomechanics and Biomedical Engineering
- Vol. 27 (6) , 736-750
Abstract
Mechanical circulatory support (MCS) devices can bridge the gap to transplant whilst awaiting a viable donor heart. The Realheart Total Artificial Heart is a novel positive-displacement MCS that generates pulsatile flow via bileaflet mechanical valves. This study developed a combined computational fluid dynamics and fluid-structure interaction (FSI) methodology for simulating positive displacement bileaflet valves. Overset meshing discretised the fluid domain, and a blended weak-strong coupling FSI algorithm was combined with variable time-stepping. Four operating conditions of relevant stroke lengths and rates were assessed. The results demonstrated this modelling strategy is stable and efficient for modelling positive-displacement artificial hearts.Keywords
Funding Information
- Scandinavian Real Heart AB
- Scandinavian Real Heart AB (107)
This publication has 32 references indexed in Scilit:
- Fluid–structure interaction, immersed boundary-finite element method simulations of bio-prosthetic heart valvesComputer Methods in Applied Mechanics and Engineering, 2013
- The Syncardia™ total artificial heart: in vivo, in vitro, and computational modeling studiesJournal of Biomechanics, 2013
- High-Resolution Fluid–Structure Interaction Simulations of Flow Through a Bi-Leaflet Mechanical Heart Valve in an Anatomic AortaAnnals of Biomedical Engineering, 2009
- Numerical Analysis on the Hemodynamics and Leaflet Dynamics in a Bileaflet Mechanical Heart Valve Using a Fluid-Structure Interaction MethodASAIO Journal, 2009
- FLUID MECHANICS OF ARTIFICIAL HEART VALVESClinical and Experimental Pharmacology and Physiology, 2009
- Curvilinear immersed boundary method for simulating fluid structure interaction with complex 3D rigid bodiesJournal of Computational Physics, 2008
- Atrioventricular plane displacement is the major contributor to left ventricular pumping in healthy adults, athletes, and patients with dilated cardiomyopathyAmerican Journal of Physiology-Heart and Circulatory Physiology, 2007
- Characteristics of Pulsatile Blood Flow Through the Curved Bileaflet Mechanical Heart Valve Installed in Two Different Types of Blood Vessels: Velocity and Pressure of Blood FlowASAIO Journal, 2006
- Validation of a Fluid–Structure Interaction Model of a Heart Valve using the Dynamic Mesh Method in FluentComputer Methods in Biomechanics and Biomedical Engineering, 2004
- Transition to turbulence in constant-mass-flux pipe flowJournal of Fluid Mechanics, 1995