Measurement of output factors for small photon beams
- 10 May 2007
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 34 (6Part1) , 1983-1988
Abstract
A variety of detectors and procedures for the measurement of small field output factors are discussed in the current literature. Different detectors with or without corrections are recommended. Correction factors are often derived by Monte Carlo methods, where the bias due to approximations in the model is difficult to judge. Over that, results appear to be contradictory in some cases. In this work, output factors were measured for field sizes from up to side length with different detectors. A simple linear correction for the energy response of solid state detectors is proposed. This led to identical values down to field size, as long as the size of the detector is small against the field size. The correction was of the order of a few percent. For a shielded silicon diode it was well below 1%. A physically meaningful function is proposed in order to calculate output factors for arbitrary field sizes with high accuracy.Keywords
This publication has 20 references indexed in Scilit:
- Application of a radiophotoluminescent glass plate dosimeter for small field dosimetryMedical Physics, 2005
- Relative output factor measurements of a 5 mm diameter radiosurgical photon beam using polymer gel dosimetryMedical Physics, 2005
- K-band EPR dosimetry: small-field beam profile determination with miniature alanine dosimeterApplied Radiation and Isotopes, 2004
- Response corrections for solid-state detectors in megavoltage photon dosimetryPhysics in Medicine & Biology, 2004
- Dosimetric properties of radiophotoluminescent glass rod detector in high‐energy photon beams from a linear accelerator and Cyber‐KnifeMedical Physics, 2004
- Comments on `Ionization chamber dosimetry of small photon fields: a Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beams'Physics in Medicine & Biology, 2003
- Ionization chamber dosimetry of small photon fields: a Monte Carlo study on stopping-power ratios for radiosurgery and IMRT beamsPhysics in Medicine & Biology, 2003
- The volume effect of detectors in the dosimetry of small fields used in IMRTMedical Physics, 2003
- Physical aspects of dynamic stereotactic radiosurgery with very small photon beams (1.5 and 3 mm in diameter)Medical Physics, 2003
- Precision dosimetry for narrow photon beams used in radiosurgery—Determination of Gamma Knife® output factorsMedical Physics, 2002