Development of a Na‐22 based pulsed slow positron beam for depth‐selective PALS

Authors:

Abstract
We developed a Na‐22 based pulsed slow positron beam system for depth‐selective positron‐annihilation lifetime spectroscopy. The use of high‐efficiency moderator and high‐compression beam bunchers enable practical measurements with using a relatively low activity positron source. An initial performance test showed timing resolution less than 250 ps and count rate of 1k counts/s with use of 0.5 GBq Na‐22.