Note: Inhibiting bottleneck corrosion in electrical calcium tests for ultra-barrier measurements
Authors:
Open Access
- 1 December 2015
- journal article
- conference paper
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 86 (12) , 126110
Abstract
A major failure mechanism is identified in electrical calcium corrosion tests for quality assessment of high-end application moisture barriers. Accelerated calcium corrosion is found at the calcium/electrode junction, leading to an electrical bottleneck. This causes test failure not related to overall calcium loss. The likely cause is a difference in electrochemical potential between the aluminum electrodes and the calcium sensor, resulting in a corrosion element. As a solution, a thin, full-area copper layer is introduced below the calcium, shifting the corrosion element to the calcium/copper junction and inhibiting bottleneck degradation. Using the copper layer improves the level of sensitivity for the water vapor transmission rate (WVTR) by over one order of magnitude. Thin-film encapsulated samples with 20 nm of atomic layer deposited alumina barriers this way exhibit WVTRs of 6 × 10−5 g(H2O)/m2/d at 38 °C, 90% relative humidity.Keywords
Funding Information
- Bundesministerium für Forschung, Technologie und Raumfahrt (03IPT602A)
This publication has 14 references indexed in Scilit:
- Breakdown and Protection of ALD Moisture Barrier Thin FilmsACS Applied Materials & Interfaces, 2015
- Experimental comparison of high-performance water vapor permeation measurement methodsOrganic Electronics, 2014
- Internal sensor compensation for increased Ca test sensitivityReview of Scientific Instruments, 2014
- Note: Influence of calcium corrosion on the performance of an adjacent permeation barrierReview of Scientific Instruments, 2014
- H2O Vapor Transmission Rate through Polyethylene Naphthalate Polymer Using the Electrical Ca TestThe Journal of Physical Chemistry A, 2013
- Permeability and corrosion in ZrO2/Al2O3 nanolaminate and Al2O3 thin films grown by atomic layer deposition on polymersJournal of Vacuum Science & Technology A, 2012
- Electrical calcium test for moisture barrier evaluation for organic devicesReview of Scientific Instruments, 2011
- Quantitative calcium resistivity based method for accurate and scalable water vapor transmission rate measurementReview of Scientific Instruments, 2011
- Evaluation of gas permeation barrier properties using electrical measurements of calcium degradationReview of Scientific Instruments, 2007