Accurate Detection of Dobutamine-induced Haemodynamic Changes by Kino-Cardiography: A Randomised Double-Blind Placebo-Controlled Validation Study
Open Access
- 19 July 2019
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 9 (1) , 1-11
Abstract
Non-invasive remote detection of cardiac and blood displacements is an important topic in cardiac telemedicine. Here we propose kino-cardiography (KCG), a non-invasive technique based on measurement of body vibrations produced by myocardial contraction and blood flow through the cardiac chambers and major vessels. KCG is based on ballistocardiography and measures 12 degrees-of-freedom (DOF) of body motion. We tested the hypothesis that KCG reliably assesses dobutamine-induced haemodynamic changes in healthy subjects. Using a randomized double-blinded placebo-controlled crossover study design, dobutamine and placebo were infused to 34 volunteers (25 ± 2 years, BMI 22 ± 2 kg/m², 18 females). Baseline recordings were followed by 3 sessions of increasing doses of dobutamine (5, 10, 20 μg/kg.min) or saline solution. During each session, stroke volume (SV) and cardiac output (CO) were determined by echocardiography and followed by a 90 s KCG recording. Measured linear accelerations and angular velocities were used to compute total Kinetic energy (iK) and power (Pmax). KCG sorted dobutamine infusion vs. placebo with 96.9% accuracy. Increases in SV and CO were correlated to iK (r = +0.71 and r = +0.8, respectively, p < 0.0001). Kino-cardiography, with 12-DOF, allows detecting dobutamine-induced haemodynamic changes with a high accuracy and present a major improvement over single axis ballistocardiography or seismocardiography.Keywords
Funding Information
- Belgian Federal Scientific Policy Office
- Fonds Erasme
This publication has 21 references indexed in Scilit:
- Biological Data Tracing and Pattern Recognition in Real-time2016
- Ballistocardiography and Seismocardiography: A Review of Recent AdvancesIEEE Journal of Biomedical and Health Informatics, 2014
- Toward Continuous, Noninvasive Assessment of Ventricular Function and Hemodynamics: Wearable BallistocardiographyIEEE Journal of Biomedical and Health Informatics, 2014
- Tracking clinical status for heart failure patients using ballistocardiography and electrocardiography signal features2014
- Myocardial contractility: A seismocardiography approach2012
- Non-invasive cardiac output trending during exercise recovery on a bathroom-scale-based ballistocardiographPhysiological Measurement, 2009
- Normal stroke volume and cardiac output response during dobutamine stress echocardiography in subjects without left ventricular wall motion abnormalitiesThe American Journal of Cardiology, 1995
- Moment of inertia of whole body using an oscillating table in adolescent boysJournal of Biomechanics, 1995
- Symptoms, adverse effects, and complications associated with dobutamine stress echocardiography. Experience in 1118 patients.Circulation, 1993
- The relation of the ballistocardiogram to cardiac functionThe American Journal of Cardiology, 1958